Chapter 5 Start-up
5. Start-up
5.1. Let the Device Adapt to the Ambient Conditions
Let the device adapt to the ambient conditions!
The device must adapt to ambient conditions prior to start-up.
• Set up the device in the operating room and let it adapt to the expected operating room
temperature.
• For a period of approx. 2 hours prior to switching it on.
• Open the device doors.
5.2. Preparing the Work Space
Upon delivery, the CO
be decontaminated.
Before the decontamination is performed, the following work space components must be cleaned:
• Support rails,
• support hooks,
• shelves,
• bottle turning assembly (optional),
• gas humidification (optional),
• work space surfaces,
• work space seals and gaskets,
• glass door/gas-tight screen.
Note
Decontamination:
For details about the cleaning and disinfection of the device (see
page 124
).
Heracell 150i / 240i GP 50159351_B_06-2020
incubator is not in a sterile state. Before the initial start-up, the device must
2
"Cleaning and Disinfection"
49
on