Physical Specifications Agilent 1260 Infinity Ii Bio Flexible Pump (G7131C) - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

2
Site Requirements and Specifications
Physical Specifications
Physical Specifications Agilent 1260 Infinity II Bio
Flexible Pump (G7131C)
Table 2
Type
Weight
Dimensions
(height × width × depth)
Line voltage
Line frequency
Power consumption
Ambient operating
temperature
Ambient non-operating
temperature
Humidity
Operating altitude
Non-operating altitude
Safety standards:
IEC, EN, CSA, UL
ISM Classification
1
At operating temperatures <15 °C the Low Temperature Environment Kit (G7131-67000) must be in-
stalled.
1260 Infinity II Flexible Pump User Manual
Physical Specifications
Specification
16.8 kg (37.0 lbs)
180 x 396 x 436 mm
(7.1 x 15.6 x 17.2 inches)
100 – 240 V~, ± 10 %
50 or 60 Hz, ± 5 %
120 VA / 110 W
4 – 55 °C (39 – 131 °F)
-40 – 70 °C (-40 – 158 °F)
< 95 % r.h. at 40 °C (104 °F)
Up to 3000 m (9842 ft)
Up to 4600 m (15092 ft)
Overvoltage category II, Pollution degree 2
ISM Group 1 Class B
Comments
Wide-ranging capability
1
See footnote
Non-condensing
For storing the module
For indoor use only.
According to CISPR 11