Solvent Handling

Handling of Normal Phase Solvents
Observe the following recommendations when using normal phase solvents:
1260 Infinity II Flexible Pump User Manual
Recommended valves for normal phase applications
Binary or High-Speed Pumps
(G7120A)
1290 Infinity Inlet Valve Type N
(G4220-60122)
1290 Infinity Outlet Valve Type N
(G4220-60128)
Always use fresh, filtered solvents. Exchange solvents every second day.
Prevent reactions caused by heat, light and oxygen. Use brown, firmly closed
bottles.
Whenever possible, use stabilizers, e.g. butylated hydroxytoluene (BHT) for
ethyl ether.
Use isopropanol (IPA) to flush out the previous solvent when converting a
system from normal phase to reverse phase, or vice versa.
Store all unused channels in IPA.
Don't turn off the pump while it is filled with solvents with low boiling points.
Generate a small flow throughout the used channels when the pump is not in
use.
Quaternary or Flexible Pumps
(G7104A/C)
1290 Infinity Quat Inlet Valve Type N
(G4204-60122)
1290 Infinity Outlet Valve Type N
(G4220-60128)