Troubleshooting and Diagnostics
Pump Leak Rate Test
Troubleshooting the Pump Leak Rate Test
Secondary Leak > 3 µL/min
•
•
Dynamic Leak > 30 µL/min or Dynamic Leak Rate Test fails
•
•
•
1260 Infinity II Flexible Pump User Manual
Leak between the OBV and automatic purge valve
•
Check for visible leaks on fittings and connectors
•
Connector not fixed / tight enough
•
Connector damaged
•
Leaky filter frit assembly
•
Remove the seal wash tubes from the support ring and check for leak into
the seal wash path
•
Main seal leaking/damaged
•
Piston damaged
•
Remove waste lines from the automatic purge valve
•
Damage to rotor seal and/or stator head
Outlet valve not properly assembled
•
Re-tighten the outlet valve
•
Check the position of the gold seal
Air in the primary pump chamber
•
Check for air in the solvent inlet lines and the Tuning signal
•
Purge the lines, condition and if necessary prime the pump head
Abort due to over pressure
•
Check solvent and solvent settings
•
Purge and condition the pump head with water
Leak in inlet valve
•
Check for moving air bubbles in tubing directly to the inlet valve
•
Purge the lines with water to remove dirt
•
Knock at the valve, clean it or replace it