7
Error Information
Pump Error Messages
Unknown multipurpose valve type
Error ID: 29242
The type information of the multipurpose valve is invalid.
•
Probable cause
1
2
Pump drive encoder error
Error ID: 29211
The pump drive encoder has generated no signal.
•
Probable cause
1
Pump drive error
Error ID: 29212, 29213
The pump drive failed during calibration.
•
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Parameter: None
Wrong valve head installed.
Valve head has invalid RFID tag content.
Parameter: 1 – 2 referring to pump drive
Pump drive encoder is defect.
Parameter: 1 – 2 referring to pump drive
Pump drive encoder is defect.
Suggested actions
Check or replace multipurpose valve head.
Check or replace multipurpose valve head.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
141