Reading The Pump Encoder Tag Failed; Writing The Pump Encoder Tag Failed - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
Pump Error Messages

Reading the pump encoder tag failed

Error ID: 29201
Reading the pump encoder tag has failed.
Probable cause
1
2

Writing the pump encoder tag failed

Error ID: 29200
Writing the pump encoder tag has failed.
Probable cause
1
2
1260 Infinity II Flexible Pump User Manual
Parameter: 1 – 2 referring to pump drive
Defect connection between encoder and
main board.
Missing or defect tag Defect connection
between tag and encoder.
Parameter: 1 – 2 referring to pump drive
Defect connection between encoder and
main board.
Defect tag.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.