Performance Specifications Agilent 1100 Series Capillary Pump

Table 86
Type
System delay volume Typically 5 µl from EFC to column head, for flow rates up to 20 µl/min (
Table 87
Type
Hydraulic system
Settable column flow
range
Recommended column
flow range
Column flow precision
Optimum composition
range
Composition precision
Table 87 Performance Specification Agilent 1100 Series Capillary Pump (continued )
1100 Series Capillary LC System Manual
Performance Specification Agilent 1100 Series Capillary LC System
Specification
default setup ).
Typically 14 µl from EFC to column head, for flow rates up to 100
µl/min ( default setup ).
Performance Specification Agilent 1100 Series Capillary Pump
Specification
Two dual piston in series, with proprietary servo-controlled variable
stroke drive, floating piston, active inlet valve, solvent selection valve
and electronic flow control for flow rates up to 100 µl/min
0.01 – 20 µl/min
0.01 – 100 µl/min (with the extended flow range kit)
0.001 – 2.5 µl/min (with the electronic flow control bypassed)
1 – 20 µl/min
10 – 100 µl/min (with extended flow range kit)
0.1 – 2.5 ml/min (with the electronic flow sensor bypassed)
< 0.7 % RSD or 0.03 % SD (typically 0.4 % RSD or 0.02 % SD), at
10 µl/min and 50 µl/min column flow (based on RT, default setting)
1 to 99% or 5 µl/min per channel (primary flow), whatever is greater
< 0.2 % SD, at 10 µl/min (20 µl flow sensor), 50 µl/min (100 µl flow
sensor) and 1 ml/min (normal mode) default setting
159

Performance Specifications

Agilent Technologies
7