7
Error Information
Pump Error Messages
Timeout: Wait for Pressure
Error ID: 29185
A target condition (pressure) has been sent to the instrument which should have
been reached within an expected time frame but didn't. Either the limit, time
frame or the current value of the variable has been modified later directly or
indirectly.
Probable cause
1
2
Drive Encoder failed
Error ID: 29210
Drive encoder failed during pump drive calibration.
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Leak
Internal error.
Suggested actions
Verify control software, macros, manual
commands.
Run system pressure test for identifying and
localizing the leak. Correct leak.
Suggested actions
Contact Agilent support.
145