8
Maintenance
Remove the Pump Head Assembly (G7104C)
Remove the Pump Head Assembly (G7104C)
Tools required
p/n
G7120-68708
1
In Lab Advisor go to Service & Diagnostics >
Remove/Install Pump Head and follow instructions given
on the screen.
3
Remove the seal wash tubes.
5
Remove the capillary connection from the outlet
adapter on the secondary pump head to the pressure
sensor.
1260 Infinity II Flexible Pump User Manual
Description
HPLC System Tool Kit-Infinity-II
2
Open the doors.
4
DO NOT REMOVE the heat exchanger connection
between the pump heads marked by the red X.
6
Open the black plastic screw of the inlet valve at the
bottom of the primary pump head.
X
176