Error Information
Pump Error Messages

Reading of multipurpose valve tag failed

Error ID: 29240
Reading the multipurpose valve tag failed.
Probable cause
1
2
3

Pump drive encoder rollover

Error ID: 29232
Invalid pump drive encoder signals have been detected.
Probable cause
1

Drive position limit

Error ID: 29234
Internal error.
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Parameter: None
Reading of multipurpose valve tag failed.
Multipurpose valve head tag defect or empty. Replace multipurpose valve head.
Multipurpose valve tag reader is defect.
Parameter: 1 – 2 referring to pump drive
Pump drive encoder is defect.
Parameter: 1 – 4 referring to pump drive
Internal error.
Suggested actions
Check cable connection.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.