Timeout: Wait For Volume; Timeout: Wait For Flow - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
Pump Error Messages

Timeout: Wait for Volume

Error ID: 29182
A target condition (volume, which is the delivered flow since the limit has been
set) has been sent to the instrument which should have been reached within an
expected time frame but didn't. Either the limit, time frame or the current value of
the variable has been modified later directly or indirectly (for example the flow
rate).
Probable cause
1

Timeout: Wait for Flow

Error ID: 29183
A target condition (flow rate) has been sent to the instrument which should have
been reached within an expected time frame but didn't. Either the limit, time
frame or the current value of the variable has been modified later directly or
indirectly.
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Suggested actions
Verify control software, macros, manual
commands.
Suggested actions
Verify control software, macros, manual
commands.
144

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents