Using the Module
Best Practices
The user-optimized Prepare Pump context menu replaces the classical menu:
1.
2.
3.
4.
1.
2.
3.
4.

Seal Wash

Seal Wash guarantees a maximum seal life time. Use Seal Wash:
Contaminated seal wash solvent
C AU T I O N
Do not recycle seal wash solvent to avoid contamination.
Weekly exchange seal wash solvent.
1260 Infinity II Flexible Pump User Manual
1290 Infinity II Pumps
Figure 12
Prepare the pump (1290 Infinity II pumps)
Right click on the module dashboard
Select Prepare Pump...
Select the procedure and fill in adequate parameters
Click Start to run the selected procedure
When using buffers with elevated salt concentrations
When using volatile solvents with non-volatile additives