7
Error Information
Pump Error Messages

Overcurrent of pump drive

Error ID: 29202
The current consumption of the pump drive is too high.
Probable cause
1
2

Deliver underrun

Error ID: 29233
Internal error.
Probable cause
1
2
1260 Infinity II Flexible Pump User Manual
Parameter: 1 – 2 referring to pump drive
Blockage of system before pressure sensor.
Drive motor defect.
Parameter: None
Internal error.
Firmware issue
Suggested actions
Check for blockage of e.g. outlet valve filter frit,
Multipurpose Valve, heat exchanger.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Use a minimum firmware revision of B.06.55
137