Timeout: Wait for Composition

Error ID: 29180
A target condition (composition) has been sent to the instrument which should
have been reached within an expected time frame but didn't. Either the limit, time
frame or the current value of the variable has been modified later directly or
indirectly.
Probable cause
1

Timeout: Wait for run volume

Error ID: 29181
A target condition (run volume, which is the volume delivered since the method
run start) has been sent to the instrument which should have been reached
within an expected time frame but didn't. Either the limit, time frame or the
current value of the variable has been modified later directly or indirectly (for
example the flow rate).
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Incorrect parameters have been sent to the
instrument by the control software or
manual changes.
Suggested actions
Verify control software, macros, manual
commands.
Suggested actions
Verify control software, macros, manual
commands.