Prepare the Pump Module for Transport

Prepare the Pump Module for Transport

When
If the module shall be transported or shipped.
Parts required
p/n
9301-0411
9301-1337
G7104-44000
G7131-98000
Mechanical damage
C AU T I O N
For shipping the module, insert the Protective Foam to protect the module
from mechanical damage.
Be careful not to damage tubing or capillary connections while inserting the
module in the Protective Foam.
1
Flush all solvent channels with isopropanol.
2
Remove solvent inlet tubes from solvent reservoirs and
tubing clips at other modules.
3
Remove tubings between the seal wash function and
solvent bottle/waste.
4
Turn off the pump.
5
Remove cable and capillary connections to other
modules.
6
Remove the waste tube.
7
Remove the module from the stack.
1260 Infinity II Flexible Pump User Manual
Description
Syringe; Plastic
Syringe adapter
Transport protection foam (G7104A/C)
Transport protection foam (G7131A/C)
8
Disconnect the degasser outlet tubings at the MCGV
one after another. Use a syringe for removing liquid
from the degasser and the solvent tubings.
229