Power Up / Shut-Down The Pump - Agilent Technologies InfinityLab LC Series User Manual

Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

3
Using the Module
Best Practices

Power up / Shut-down the pump

Prepare the Pump
Shut Down the System
1260 Infinity II Flexible Pump User Manual
Use fresh or different mobile phase (as required).
Purge each channel with 2.5 – 3 mL/min for 5 min. Open the manual purge
valve or use the purge command, depending on the pump type.
Flush the column with the appropriate solvents and store it according to
column manual instructions (be sure that the flushing solvent is compatible
with the solvent present in the system to avoid precipitation).
Install a union or a restriction capillary and flush the system extensively with
water, especially after using buffers. For details, see
page 56.
Flush and store the system in 50 % methanol or 50 % isopropanol in water,
without additives.
Remove all samples from the autosampler and store them according to good
laboratory practices.
Power off all modules.
"Flushing Procedure"
on
43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents