8
Maintenance

Replace Parts of the Multipurpose Valve (G7131C)

Replace Parts of the Multipurpose Valve
(G7131C)
Tools required
p/n
8710-2394
Parts required
p/n
5068-0379
5068-0378
5428-0006
Preparations
Remove all capillary connections from the Multipurpose Valve.
1 Use the 9/64 inch hex key for opening the valve head.
2 Replace parts as required.
3 Reassemble the valve head mounted to the valve drive.
Rotor seal
Stator head
Stator screws
1260 Infinity II Flexible Pump User Manual
Description
9/64 inch hex key
Description
Rotor Seal Bio-Compatible, Multipurpose Valve Bio-Compatible, PEEK, 1300 bar
Stator head Bio-Compatible
Stator screws Bio-Compatible