Error Information
Pump Error Messages
Flow rate limit exceeded
Error ID: 29164
The flow rate of the quaternary pump has exceeded the limit, while the pump
runs in pressure controlled mode, e.g. during a pressure test.
•
Probable cause
1
2
3
4
Quaternary pump shutdown during analysis
Error ID: 29199
The quaternary pump has been shut down by the control software or control
module during an analysis.
•
Probable cause
1
1260 Infinity II Flexible Pump User Manual
Parameter: None
Leak
Bottle empty.
Shutoff valve closed (if applicable).
Drift of pressure sensor (unlikely for short
tests taking some minutes).
Parameter: 0 for off, 1 for standby.
Pump has been shut down.
Suggested actions
Check for leaks in the pump and flow path.
Fill solvent bottle.
Open shutoff valve.
Replace pressure sensor.
Suggested actions
Restart pump.