8
Maintenance

Replace the Multipurpose Valve

Replace the Multipurpose Valve

Tools required
p/n
5023-2502
Parts required
#
1
1
1
2
1
1
1
1
1
Remove all capillary connections from the Multipurpose
Valve.
3
Unscrew the black union nut and remove the head of
the purge valve by pulling it to the front.
1260 Infinity II Flexible Pump User Manual
Description
Hex driver SW-6.35, slitted
p/n
Description
0100-1259
Blank nut (plastic)
01080-83202
Blank nut (stainless steel)
5067-4237
Multipurpose Valve Head
5500-1253
Capillary ST 0.17 mm x 130 mm SX/S (OPTIONAL)
5067-6704
Multipurpose Valve Bio-Compatible
5043-0277
Blank nut long 10-32, PEEK with stainless steel core
5500-1284
Capillary MP35N 0.17 mm x 120 mm SI/SX
5004-0041
Capillary MP35N 0.17 mm x 130 mm SI/SX
2
Remove the clamp with the inline filter (if installed).
4
Put the new valve head onto the valve drive such that
the lobe fits to the groove. Screw the valve head onto
the valve drive using the union nut.
215