Maintenance

Replace the Outlet Valve (G7131C)

Replace the Outlet Valve (G7131C)

When
If Outlet valve is defective.
Tools required
p/n
8710-0510
Parts required
p/n
G7131-60028
Preparations
1
Close the shut off valves to avoid solvent leaks.
1260 Infinity II Flexible Pump User Manual
Description
Open-end wrench 1/4 — 5/16 inch
Spanner-double open ended 14 mm
Description
Outlet Valve Bio-Compatible
Switch off pump at the main power switch
Open the doors
Use an optional solvent shutoff valve or lift up solvent filters inside solvent bottles for avoiding
leakages
Remove the pump head from the module
2
Open the lock screw of the heat exchanger capillary.