Owon XDS4000 Series User Manual page 98

Digital storage oscilloscopes
Hide thumbs Also See for XDS4000 Series:
Table of Contents

Advertisement

Line
Time
Type
Voltage
(Time&Volt
age type)
Window
Main
(Wave
zoom
Extension
mode)
a
b
Line
ab
Source
CH1 to CH4
Perform the following operation steps for the time and voltage cursor
measurement of the channel CH1:
1. Push Cursor to display the cursor menu.
2. Select Source in the bottom menu, select CH1 in the right menu.
3. Select the first menu item in the bottom menu, the Type menu will
display at the right of the screen. In the right menu, select
Time&Voltage for Type, two blue dotted lines displayed along the
horizontal direction of the screen, two blue dotted lines displayed along
the vertical direction of the screen. Cursor measure window at the left
bottom of the screen shows the cursor readout.
4. In the bottom menu, select Line Type as Time to make the vertical
cursors active. If the Line in the bottom menu is select as a, turn the M
knob to move line a to the right or left. If b is selected, turn the M knob
to move line b.
5. In the bottom menu, select Line Type as Voltage to make the
horizontal cursors active. Select Line in the bottom menu as a or b,
turn the M knob to move it.
6. Push the Horizontal HOR button to enter wave zoom mode. In the
bottom cursor menu, select Window as Main or Extension to make
the cursors shown in the main window or zoom window.
4.Advanced User Guidebook
The
horizontal
intersections of the vertical cursors and the
waveform
Makes the vertical cursors active.
Makes the horizontal cursors active.
Measure in the main window.
Measure in the extension window.
Turn the M knob to move line a.
Turn the M knob to move line b.
Two cursors are linked. Turn the M knob to
move the pair of cursors.
Display the channel to which the cursor
measurement will be applied.
92
cursors
are
set
as