7.Communication with PC
Figure 7-6 Set the network parameters of the Oscilloscope Software
(4) Set the network parameters of the oscilloscope. In the oscilloscope, push
the Utility button. Select Function in the bottom menu. Select LAN Set in the
left menu. In the bottom menu, set the Type item as LAN, and select Set. In
the right menu, set IP and Port to the same value as the "Ports-settings" in the
software in step (3). The Netgate and Net mask should be set according to the
router. Select Save set in the bottom menu, it prompts "Reset to update the
config". After resetting the oscilloscope, if you can get data normally in the
oscilloscope software, the connection is successful.
Figure 7-7 Set the network parameters of the oscilloscope