Owon XDS4000 Series User Manual page 57

Digital storage oscilloscopes
Hide thumbs Also See for XDS4000 Series:
Table of Contents

Advertisement

Chk Error
Mode
Normal
Holdoff
2. I2C Trigger
The I2C serial bus consists of SCL and SDA. The transmission rate is
determined by SCL, and the transmission data is determined by SDA. As
shown in below figure, oscilloscope can trigger on the start, restart, stop, ack
lost, specific device address or data value, also device address and data
value at the same time.
In I2C bus trigger mode, the trigger setting information is displayed on
bottom right of the screen, for example,
trigger type is I2C, CH1 trigger level is 0.00mV.
I2C Trigger menu list:
MENU
SETTING
Bus Type
Source
When
4.Advanced User Guidebook
Even-Odd:Select Even or Odd.
Common Baud: adjust M knob to choose
common baud.
Custom Baud: adjust M knob to choose
baud, ranges from 50 to 10,000,000.
Data Bits:Set as 5、6、7、8 bits.
Data
Data:Set data according to data bits,
ranges from 0-31, 0-63, 0-127 or 0-255.
Auto
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
Single
When trigger occurs, acquire one
waveform then stop
INSTRUCTION
I2C
Set vertical channel bus type as I2C trigger.
SCL
Set SCL.
SDA
Set SDA.
Trigger when SDA data transitions from
Start
high to low while SCL is high.
When another start condition occurs before
Restart
a stop condition.
51
, indicates that