Owon XDS4000 Series User Manual page 49

Digital storage oscilloscopes
Hide thumbs Also See for XDS4000 Series:
Table of Contents

Advertisement

Auto
Normal
Single
Mode
Holdoff
Holdoff
5.Runt Trigger
Trigger pulses that pass through one trigger level but fail to pass through the
other trigger level. Shown as below figure,
In Runt Trigger mode, the trigger setting information is displayed on bottom
right of the screen, for example,
type is runt, trigger source is CH1, polarity is positive, 0.00mV is the differential
between up level and low level threshold.
Runt Trigger menu list:
MENU
SETTING
Single
Runt
Mode
CH1
CH2
Source
CH3
CH4
4.Advanced User Guidebook
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform then
stop
100 ns – 10 s, turn the M knob or click
time interval before another trigger occur, press
panel button or click
cursor to choose which digit to be set.
Runt Trigger
INSTRUCTION
Set vertical channel trigger type as runt trigger.
Select CH1 as the trigger source
Select CH2 as the trigger source
Select CH3 as the trigger source
Select CH4 as the trigger source
43
to set
to move
, indicates that trigger