User's Manual of OWON Color Mixed Signal Digital Storage Oscilloscope
OWON
MSO5022
Portable Mixed Signal Digital Storage Oscilloscope
User Manual
WWW.OWON.COM.UA
1

Advertisement

Table of Contents
loading

Summary of Contents for Owon MSO5022

  • Page 1 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope OWON MSO5022 Portable Mixed Signal Digital Storage Oscilloscope User Manual WWW.OWON.COM.UA...
  • Page 2 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Copy Right in this Manual © Lilliput Company. All rights have been reserved. The Lilliput’s products are under the protection of the patent rights in America and other countries, including ones which have already obtained the patent rights and those which are applying for.
  • Page 3 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope WARRANTY Lilliput warrants that this OWON brand product will be free from defects in materials and workmanship for a period of three years (one year limited warranty applies to the accessories) from the date of shipment.
  • Page 4: Table Of Contents

    Table of Contents General Safety Requirements....................6 Safety Terms and Symbols....................7 General Characteristics of the MSO5022 Color Mixed Signal Digital Storage Oscilloscope ......................... 9 Junior User Guidebook ..................... 10 Introduction to the Front Panel and the User's Interface of the MSO series Oscilloscope ..11 Front panel .............................
  • Page 5 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Do Self Cal (Self-Calibration)........................ 57 SYS STAT (System State) ........................57 How to Conduct the Automatic Measurement ..................58 How to Carry on the Cursor Measurement..................... 60 Cursor Measurement ..........................61 AUTOSET..............................
  • Page 6: General Safety Requirements

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope General Safety Requirements Before any operations, please read the following safety precautions to avoid any possible bodily injury and prevent this product or any other products connected from damage. In order to avoid any contingent danger, this product is only used within the range specified.
  • Page 7: Safety Terms And Symbols

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Safety Terms and Symbols Terms in this manual. The following terms may appear in this manual: Warning. A warning statement indicates the conditions and actions which may endanger the life safety.
  • Page 8 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope To avoid body damage and prevent product and connected equipment dam. This product can only be used in the specified applications. Carefully read the following safety information before using the test tool.
  • Page 9: General Characteristics Of The Mso5022 Color Mixed Signal Digital Storage Oscilloscope

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope General Characteristics of the MSO5022 Color Mixed Signal Digital Storage Oscilloscope Digital Storage Oscilloscope With the bandwidth of 25M; Record length of 5,000 points for each channel; Sampling rates of 100MS/s for each channel;...
  • Page 10: Junior User Guidebook

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Junior User Guidebook This chapter deals with the following topics mainly: Digital Storage Oscilloscope Introduction to the front panel and the user’s interface of the MSO series oscilloscope How to carry on the general inspection...
  • Page 11: Introduction To The Front Panel And The User's Interface Of The Mso Series Oscilloscope

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Introduction to the Front Panel and the User's Interface of the MSO series Oscilloscope When you get a new-type oscilloscope, you should get acquainted with its front panel at first and the MSO series mixed digital storage oscilloscope is no exception. This chapter...
  • Page 12 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 6、 DSO signal input 7、 Measurement signal output 8、 Power and charging indication: Green light indicate AC supply and battery full charged; yellow light indicate under charging Fig 2: Keys Overview 1、Menu option setting: F1~F5...
  • Page 13 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope position of CH1. CH2, and two “Volts/Div” knob control voltage scale of CH1, CH2. For LA, “CH1 menu”, “CH2 menu”, “Wave math” keys and “CH2 Volts/Div” knob are idle. “CH1 Vertical”, “CH2 Vertical” to adjust the M1, M2 position in Cursor menu when cursor display is on.
  • Page 14: Digital Storage Oscilloscope

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Digital Storage Oscilloscope User interface introduction Fig. 3 Illustrative Drawing of Display Interfaces 1. The Trigger State indicates the following information: Auto: The oscilloscope is under the Automatic mode and is collecting the waveform under the non-trigger state.
  • Page 15 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 5. This reading shows the time deviation between the horizontal trigger position and the window centre line, which is regarded as 0 in the window center. 6. It indicates the current function menu.
  • Page 16: How To Carry On The General Inspection

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 22. The positions of two purple dotted line cursors measurements. How to Carry on the General Inspection After you get a new MSO series oscilloscope, it is recommended that you should make a check on the instrument according to the following steps: 1.
  • Page 17: How To Implement The Probe Compensation

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Connect the probe tip and the ground clamp to the connector of the probe compensator,. 3. Press the “7(AUTOSET)” Button. The square wave of 1 KHz frequency and 5V peak-peak value will be displayed in several seconds (see Fig.
  • Page 18: How To Set The Probe Attenuation Coefficient

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5 Displayed Wave Forms of the Probe Compensation 3. Repeat the steps mentioned if necessary. Fig. 6 Adjust Probe How to Set the Probe Attenuation Coefficient The probe has several attenuation coefficients, which will influence the vertical scale factor of the oscilloscope.
  • Page 19: How To Use The Probe Safely

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 7 Attenuation Switch Note: When the attenuation switch is set to 1X, the T5100 probe will limit the bandwidth of the oscilloscope in 5MHz. If it is needed to use the whole bandwidth of the oscilloscope, the switch must be set to 10X.
  • Page 20: Introduction To The Vertical System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope “ Do Self Cal”; finally, run the program after confirming that everything is ready now. Introduction to the Vertical System Shown as Fig.9, there are a series of buttons and knobs in VERTICAL CONTROLS.
  • Page 21: Introduction To The Horizontal System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope corresponding to the status bar has been changed accordingly. Press buttons of “CH1 MENU”, “CH2 MENU” and “MATH MENU”, the operation menu, symbols, wave forms and scale factor status information of the corresponding channel will be displayed in the screen.
  • Page 22: Introduction To The Trigger System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Introduction to the Trigger System Shown as Fig.11, there are a knob and four buttons in the “TRIGGER CONTROLS”. The following practices will direct you to be familiar with the setting of the trigger system gradually.
  • Page 23: Logic Analyzer

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Logic Analyzer LA input connection Insert the plug of OL-16 LA module 50P into the LA signal input on front panel and fix two screw. Then 16 channel clamp of OL-16 LA connect to target...
  • Page 24: How To Acquire Data

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope different display。 11、Sample status indicate: “RUN” for sampling and wait for trigger, “TRIG” for trigger detected and wait for sample finished. “STOP” for sampling finished。 12、Value indicate current time base。...
  • Page 25: Display Systems

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope We will use a simple measurement example to explain the primary setting for LA measurement. We need to measure a three lines SPI signal, three signal are enable, clock and data.
  • Page 26: Trigger System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Figure 14 Figure 15 Trigger system LA is same as DSO and need to make trigger to synchronize data. The trigger system mainly to set trigger sources, trigger mode and trigger position.
  • Page 27: Threshold Voltage System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Then trigger system setting finished (ref to fig. 16). Figure 16 Threshold voltage system Threshold voltage system is to set high/low of the trigger voltage. The system already fixed the setting for normal logic voltage as CMOS, LVMOS etc. And you can set any trigger voltage using custom setting.
  • Page 28 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope in the memory. The recorded data will display in error if the sample rate is too lower. Below figures explains how sample rate influence the waveform recorded in LA. Figure 18 There is an importance compromise between recorded signal resolution and its continuance (relate to time).
  • Page 29: Advanced User Guidebook

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Advanced User Guidebook Up till now, you have already been familiar with the initial operations of the functions of the function areas, buttons and knobs in the front panel of the MSO series oscilloscope.
  • Page 30: Digital Storage Oscilloscope

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope It is recommended that you read this chapter carefully to get acquainted the various measurement functions and other operation methods of the MSO series oscilloscope. Digital Storage Oscilloscope How to Set the Vertical System...
  • Page 31 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The description of the Channel Menu is shown as the following list: following form: Function Menu Setting Description Coupling Block the AC component in the input signal. Unblock the AC and DC components in the input signal.
  • Page 32 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 21 AC Coupling Oscillogram Fig. 22 DC Coupling Oscillogram...
  • Page 33 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Setting the Channel “ON/OFF” Taking the Channel 1 for example, the operation steps are shown as below: (1). Press the CH1 MENU button and call out the CH1 SETUP menu.
  • Page 34 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope A List of the Attenuation Coefficient of Probes and the Corresponding Menu Settings. Attenuation Coefficient of the Probe Corresponding Menu Setting 10:1 100:1 100X 1000:1 1000X Setting of Wave Form Inverted Wave form inverted: the displayed signal is turned 180 degrees against the phase of the earth potential.
  • Page 35 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 24 Wave Form not inverted Fig. 25 Wave Form Inverted...
  • Page 36: Implementation Of Mathematical Manipulation Function

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Implementation of Mathematical Manipulation Function The Mathematical Manipulation function is used to show the results of the additive and subtraction operations between Channel 1 and Channel 2. Taking the additive operation between Channel 1 and Channels 2 for example, the operation steps are as follows: 1.
  • Page 37: Application Of Vertical Position And Volts/Div Knobs

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Application of VERTICAL POSITION and VOLTS/DIV Knobs 1. The .VERTIVAL POSITION knob is used to adjust the vertical positions of the wave forms of all Channels (including those resulted from the mathematical operation).
  • Page 38: How To Set The Horizontal System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to Set the Horizontal system The HORIZONTAL CONTROLS includes the HORIZONTAL NENU button and such knobs as HORIZONTAL POSITION and SEC/DIV. 1. HORIZONTAL POSITION knob: this knob is used to adjust the horizontal positions of all channels (include those obtained from the mathematical manipulation), the analytic resolution of which changes with the time base.
  • Page 39: Window Setting

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Press the F1 menu selection button and choose the Main Time Base. In this case, the HORIZONTAL POSITION and SEC/DIV knobs are used to adjust the main window. The display in the screen is shown as Fig. 29.
  • Page 40: Window Expansion

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 30 Window Setting Window Expansion Press the F3 menu selection button and choose Zone Window. As a result, the window area defined by two cursors will be expanded to the full screen size (see Fig.
  • Page 41: How To Set Trigger System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to set trigger system Trigger is to determine when DSO starts to acquire data and waveform display. Once trigger to be set correctly then it will convert the unstable display to meaning waveform.
  • Page 42 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 33 Edge trigger menu Edge menu list: MENU SETTING INSTRUCTION Slop Rising Trigger in signal rising edge Falling Trigger in signal falling edge Source Select CH1 as the trigger source.
  • Page 43 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Follow up below step to set CH1 as rising slope edge trigger and trigger mode is auto, coupling is DC, trigger sensitivity is 0.5 Div. 1、Press “Trig menu” 2、Press “F1” to choose “Edge”...
  • Page 44 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 图 Fig. 35 Edge trigger (Falling Wave) Video Choose “Video” and trigger in field/line of NTSC, PAL standard video signal Trig menu refer to Fig.36 Fig. 36 Video trigger menu...
  • Page 45 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Video trigger menu MENU SETTING INSTRUCTION POLARITY NORMAL For low video signal of black voltage level REVERSE For high video signal of black voltage level SOURCE Select CH1 as the trigger source.
  • Page 46 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 38 Video line trigger Oscillogram Alternate trigger Trigger signal come from two vertical channels with alternate trigger. This mode is to observe two unrelated signals. It is able to set different trigger mode (edge or video) for different channels in the menu.
  • Page 47: How To Operate The Function Menu

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope setting same as video. CH1 to measure a square wave of 10K Hz and CH2 to measure a video signal. The alternate setting as below: 1、Press “Trig menu” 2、Press “F1” and choose type as “Alternate”...
  • Page 48: How To Implement Sampling Setup

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope CURSOR, DISPLAY, AUTOSET, RUN/STOP and HARDCOPY. How to Implement Sampling Setup Press the ACQUIRE button and the menu is displayed in the screen, shown as Fig. Fig. 41 ACQU MODE Menu...
  • Page 49 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 42 Peak Detect mode, under which the burrs on the falling edge of the square wave, can be detected and the noise is heavy. Fig. 43 Common ACQU Mode display, in which no burr can be detected.
  • Page 50: How To Set The Display System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 44 The displayed wave form after the noise is removed under the Average Mode, in which the average number of 64 is set. How to Set the Display System Push down the DISPLAY button and the menu displayed in the screen is shown as Fig.
  • Page 51 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The description of the Display Set Menu is shown as follows: Function Menu Setting Description Type Vectors The space between the adjacent sampling points in the display is filled with the vector form.
  • Page 52 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 46 Display in the Vector Form Fig. 47 Display in Dots form...
  • Page 53 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Persist When the Persist function is used, the persistence display effect of the picture tube oscilloscope can be simulated: the reserved original data is displayed in fade color and the new data is in bright color. With the F2 menu selection button, different persistence time can be chosen: 1sec, 2sec, 5sec, Infinite and Closed.
  • Page 54: How To Save And Recall A Wave Form

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Auto Set Time base control Trigger control Operation steps: 1. Press the DISPLAY button and call out the Display Set Menu. 2. Press the F3 menu selection button and choose the form as XY. The display format is changed to be XY mode (see Fig.
  • Page 55: Save And Recall The Wave Form

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 50 Wave Form Save Menu The description of the Wave Form Save Menu is shown as the following table: Function Menu Setting Description Source Choose the wave form to be saved.
  • Page 56: How To Carry On The Auxiliary System Function Setting

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 51 Wave Saving How to Carry on the Auxiliary System Function Setting Press the UNTILITY button and the menu is displayed in the screen as Fig. 52. Fig. 52 Function Menu...
  • Page 57: Do Self Cal (Self-Calibration)

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The description of the Auxiliary Function Menu is shown as the following table. Function Menu Setting Description System Status Display the system function menu. Recall Factory Call out the factory settings.
  • Page 58: How To Conduct The Automatic Measurement

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The “SYS STAT” menu is described as the following table: Function Menu Setting Description Horizontal Show the horizontal parameter of the channel. Vertical Show the vertical parameter of the channel.
  • Page 59 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 55 Measure Menu Measure Four automatic measured values can be detected on each channel simultaneously. Only if the waveform channel is in the ON state can the measurement be carried out.
  • Page 60: How To Carry On The Cursor Measurement

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 56 Automatic Measurement How to Carry on the Cursor Measurement Press the CURSOR button to display the cursor measurement function menu (CURS MEAS) in the screen, which includes Voltage Measurement and Time Measurement, shown as Fig.
  • Page 61: Cursor Measurement

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The description of the cursor measurement menu is shown as the following table: Function Menu Setting Description Type Switch off the cursor measurement. Voltage Display the voltage measurement cursor and menu.
  • Page 62 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 58 Waveform of Voltage Cursor Measurement Carry out the following operation steps for the time cursor measurement of the channel CH1: 1. Press “CURSOR” and recall the CURS MEAS menu.
  • Page 63: Autoset

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 59 Wave Form of Cursor Measurement AUTOSET This button is used for the automatic setting of all control values of the instrument to generate the waveform suitable for observation. Press the AUTOSET button and the oscilloscope will perform the fast automatic measurement of the signal.
  • Page 64: Run/Stop

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope RUN/STOP Enable or disable the waveform sampling. Notice: Under the Stop state, the vertical division and the horizontal time base of the waveform can be adjusted within a certain range, in other words, the signal can be expanded in the horizontal or vertical direction.
  • Page 65: Logic Analyzer

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Logic analyzer How to set sampling system Sampling system is to set sample rate, storage depth and filter. Different sampling setting will result in different measure results. In the same storage depth, the higher sample rate set, the shorter the continuance time for signal will be.
  • Page 66 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Sample Storage depth Continuance Sample Storage depth Continuance rate time rate time 100MHz Low memory 160 us 20kHz Low memory 800 ms General 2.560 ms General 12.8 s Deep memory...
  • Page 67: How To Set Trigger System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Then sample system setting finished (refer to fig 61). Fig 61: Sampling setting How to set trigger system Trigger system including “triger level ”adjust knob, “trigger menu”, “SET 50%”, “Set to Zero”. Trig level adjust knob is to adjust trigger position. Since the digital storage depth is fixed so the proportion of storage data between pre-trigger points and re-trigger ones depend on trigger position.
  • Page 68 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Edge trigger function menu as below: Function Settings Instructions Signal choice CH00~CH0F CH00-CH0F can be set as trigger resource Slope Rising Trigger on the rising edge. Falling Trigger on the falling edge.
  • Page 69 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig 64:Bus trigger menu Bus trigger menu as below: Function Setting Instruction Source BUS0~BUS3 BUS0 to BUS3 can be set as trigger source Code 0x0000~0xffff Can be set discretionarily between 0x0000 and 0xffff (HEX)...
  • Page 70 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope position display as “NEXT T POS = 50%”. BUS trigger setting finished and ready for data acquisition (ref to Fig. 65). Fig 65:Bus trigger 3. Pattern trigger: Set channel as signal source and make high/low voltage for channel as trigger condition combination to get trigger (refer to Fig.
  • Page 71 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Pattern trigger menu as below: Function Setting Instruction Channel CH00~CH0F Select the channel to set signal pattern. choice x0100000 16 channel status indicator. 00000000 X:don’t care 0:low 1:high CH00~CH0F Don’t Care No trigger occur.
  • Page 72 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig 68:Sequential queue trigger Sequential queue trigger function as below: Function Setting Instruction Source BUS0~BUS3 Select the trigger source from BUS0~BUS3 Code 0x0000~0xffff Can be set discretionarily between 0x0000 and 0xffff (HEX)...
  • Page 73 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope again. H. Turn “Trigger level ” adjust knobtill or press “set 50%”till trigger position display as “NEXT T POS = 50%”. Then sequential queue trigger setting finished (refer to Fig 69) and data ready for acquisition.
  • Page 74 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Distributed queue trigger function as below: Function Setting Instruction Source BUS0~BUS3 Select the trigger source from BUS0~BUS3 Code 0x0000~0xffff Can be set discretionarily between 0x0000 and 0xffff (HEX)or between 0 and 65535 (DEC)...
  • Page 75 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Duration trigger function as below: Function Setting Instruction Source BUS0~BUS3 Select the trigger source from BUS0~BUS3 Code 0x0000~0xffff Can be set discretionarily between 0x0000 and 0xffff (HEX)or between 0 and 65535 (DEC)...
  • Page 76: How To Set Threshold

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig 73:Data width trigger How to set threshold Threshold setting is quite important because wrong setting will result in wrong measurement. For example, if measure signal is LVCMOS1.8V and set threshold as “CMOS/ (...
  • Page 77 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Function Setting Instruction CH SEL CH00~CH03 16 channels can be divided into 4 groups to CH04~CH07 have individual setting CH08~CH0B CH0C~CH0F Threshold CMOS/(2.5V) CMOS level and set threshold voltage as 2.5V voltage LVCMOS3.3/(1.7V)
  • Page 78: How To Set Display System

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to set display system Display system is to set on/off for channel and BUS, also to adjust the contrast of panel display. Press “A(DISPLAY)” and panel display as Fig. 76 Fig 76:Display menu...
  • Page 79: How To Set Bus

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope steps 5.6 and set BUS2, BUS3 as “OFF”. 7、Press “F1” till signal source display as “Channel” 8、Press “F2” or turn “CH1 Volts/div” knob till channel display as “CH00”. 9、Press “F3” to choose signal source as “ON”.
  • Page 80 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig 78:Bus menu BUS setting menu as below: Function Setting Instruction BUS0~BUS3 Choose BUS for operating Channel CH0F~CH00 Choose any channel among CH00~CH0F 1X111111 BUS channel complex indication: 1 for include; X...
  • Page 81: How To Measure

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig 79:Bus setting How to measure Measurement can take auto measure for values of 4 BUS synchronously. Press “measure” and BUS value for current cursor position will display directly in measurement window.
  • Page 82: How To Save And Recall

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to save and recall Use the storage menu to save or recall waveforms and setting. The storage depth of waveform for sampling storage is normal (256K) and low storage (16K).
  • Page 83: How To Use Usb Mass Storage Device To Storage

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 5. Press “F1” till memory SEL as“setups” 6. Press “F2” till memory ID. display as “setups 0” 7. Press “F3” to save current various settings. Info window show “setting saved successfully”.
  • Page 84 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig 83:Search trigger position 2. Search specified value in BUS Searching BUS menu as below: Function Setting Instruction BUS0~BUS3 Choose searching BUS Code type 0x0000~0xffff Can be set discretionarily between 0x0000 and 0xffff (HEX)or between 0 and 65535...
  • Page 85 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope “Got the target” and if it hasn’t it will show “Search failed”. G、 Press “F5” and choose “ next” to searching the matched value after current cursor. Fig 84:Search Bus 3.
  • Page 86 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope “CH00” D、Press “F3” and set CH00 as “high”. Then CH00 in “CH Sel” refer to “1” E 、 Repeat the operation of steps C. D and set other channels status corresponding to “00000000000010X1”...
  • Page 87: How To Review Setting Info

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to review setting info You can choose system information display to be “On” or “OFF” by press “INFO” key. System information including all settings for acquired waveform and next acquisition.
  • Page 88 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Cursor measurement time menu as below: Function Setting Instruction Increment Time The time difference between two cursors. Frequency The frequency difference between two cursors. Cursor 1 Time The time display of cursor 1 corresponds to trigger position.
  • Page 89 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Cursor measure position menu as below: Function Display Instruction M1-M2 Position The position difference between two cursors in memory area. Cursor 1 Position The position of cursor 1 corresponds to trigger in memory area.
  • Page 90: How To Set Utility

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to set Utility Utility function including recall factory, Language, Carry. Utility menu as below Function Setting Instruction Recall Default setting for LA factory Language Chinese Default to be Chinese...
  • Page 91: Demonstration

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Demonstration Example 1: Measurement of Simple Signals Observe an unknown signal in the circuit, and display and measure rapidly the frequency and peak-to-peak value of the signal. Carry out the following operation steps for the rapid display of this signal: 1.
  • Page 92: Example 2: Gain Of The Amplifier In The Metering Circuit

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 90 Waveform of Automation Measurement Example 2: Gain of the Amplifier in the Metering Circuit Set the probe menu attenuation coefficient as 10X and that of the switch in the probe as 10X.
  • Page 93: Example 3: Capture The Single Signal

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 7. Press the F2 menu selection button and choose Pk-Pk. 8. Press the F3 menu selection button and choose Pk-Pk. 10. Read the peak-to-peak values of Channel 1 and Channel 2 from the displayed menu (see Fig.
  • Page 94 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope level, the trigger level should be set to 2 volts and the trigger edge be set as the rising edge trigger. If it is uncertain as to the signal, you can make an observation of it in advance under the automatic or ordinary mode to determine the trigger level and the trigger edge.
  • Page 95: Example 4: Analyze The Details Of A Signal

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 92 Capture the Single Signal Example 4: Analyze the Details of a Signal Observe the Signal Containing Noises If the signal is interfered by the noise, the noise may cause a failure in the circuit. For...
  • Page 96 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 93 Wave Form of the Signal Containing Noises Separate Noises from the Signal When analyze the wave form of a signal, you should remove the noise contained in it.
  • Page 97: Example 5: Application Of X-Y Function

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 94 Wave Form of the Noise-Removed Signal Example 5: Application of X-Y Function Examine the Phase Difference between Signals of two Channels Example: Test the phase change of the signal after it passes through a circuit network.
  • Page 98: Example 6: Video Signal Trigger

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 6. Push down the DISPLAY button and recall the DISP SET menu. 7. Press the F3 menu selection button and choose XY for Format. The oscilloscope will display the input and terminal characteristics of the network in the Lissajous graph form.
  • Page 99 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope video output signal display. Video Field Trigger For the trigger in the video field, carry out operations according to the following steps: 1. Press the TRIG MENU button to display the trigger menu.
  • Page 100 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 4. Press the F3 menu selection button and choose CH1 for Source. 5. Press the F4 menu selection button and choose Line for Sync. 6. Adjust the VOLTS/DIV, VERTICAL POSITION and SEC/DIV knobs to obtain the proper wave form display (see Fig.
  • Page 101: 101

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope F.A.Q 1. In the case of that the oscilloscope is still in the black-screen state without any display after the power is switch on, implement the following fault treatment procedure.
  • Page 102: Appendix A: Technical Specifications

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Appendix A: Technical Specifications Unless otherwise specified, the technical specifications applied are applicable to the probe with the attenuation switch setting 10X and the MSO series digital oscilloscope. Only if the oscilloscope fulfill the following two conditions at first, can these specification standards be reached.
  • Page 103 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Vertical A/D Converter Resolution of 8 bits, with the sampling carried out in two channels at the same time. Sensitivity (Volt/division) 5mV/div- 5V/div (at the input BNC) Range (V/div ) Displacement Range ±...
  • Page 104: Logic Analyzer

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Trigger Level Inside ± 0.3 div Accuracy (Typical): It ± (40mV + 6% of Set Value) is applicable to the EXT/5 ± (200mV +6% of Set Value) signal with the rising and falling time ≥...
  • Page 105 User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Threshold level 0~4V Input signal range 0~5V Trigger position Pre-trigger, mid-trigger, re-trigger setting Trigger Mode Bus trigger, state trigger, data alignment trigger, data width trigger Data Search Support Data System...
  • Page 106: Appendix B: Enclosure

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Dimension 350mm× 157mm×120mm Weight 1 kilogram Interval Period of Adjustment: One year is recommended for the calibration interval period. Appendix B: Enclosure Standard Accessories: Passive probe: 2, 1.2 m, 1:1 (10:1) Probe adjustment tools CD: 1, for operation instruction.
  • Page 107: Appendix D: Battery Using Guide

    User’s Manual of OWON Color Mixed Signal Digital Storage Oscilloscope chemical cleaning agent to prevent the instrument or probe from damage. Warn: Before power on again for operation, it is required to confirm that the instrument has already been dried completely, avoiding any electrical short circuit or bodily injury resulting form the moisture.