Page 1
PDS Series Portable Mixed Signal Digital Storage Oscilloscope User Manual ■PDS8102T ■PDS8202T WWW.OWON.COM.CN...
Page 2
OWON is the registered trademark of the Lilliput Company. Xiamen Lilliput Technology Co.,Ltd.: the 5th floor, B Area, Chuangxin Mansion, Software Park,ZhenZhuWan, Huandao RD, Xiamen,Fujian,China Fujian Lilliput Optoelectronics Technology Co.,Ltd.:The mansion of optoelectronics hengsan road,...
Page 3
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope General Warranty The Lilliput warrants that the product will be free from defects in materials and workmanship for a period of three years from the date of purchase of the product by the original purchaser from the Lilliput company.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Table of Contents 1.General Safety Requirements................... 3 2.Safety Terms and Symbols....................4 3.General Characteristics of the PDS Series Oscilloscope ..........6 4.Junior User Guidebook ....................7 Introduction to the Front Panel and the User's Interface of the PDS series Oscilloscope..8 Front panel ...............................
Page 5
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 6.Demonstration ........................ 68 Example 1: Measurement of Simple Signals................68 Example 2: Gain of the Amplifier in the Metering Circuit ........... 69 Example 3: Capture the Single Signal ..................70 Example 4: Analyze the Details of a Signal................
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 1.General Safety Requirements Before any operations, please read the following safety precautions to avoid any possible bodily injury and prevent this product or any other products connected from damage. In order to avoid any contingent danger, this product is only used within the range specified.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 2.Safety Terms and Symbols Terms in this manual. The following terms may appear in this manual: Warning. A warning statement indicates the conditions and actions which may endanger the life safety.
Page 8
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope To avoid body damage and prevent product and connected equipment dam. This product can only be used in the specified applications. Carefully read the following safety information before using the test tool.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 3.General Characteristics of the PDS Series Oscilloscope Digital Storage Oscilloscope Model Bandwidth Sample Rate PDS8102T 100MHz 1GS/s half channel*, 500MS/s each channel PDS8202T 200MHz 2GS/s half channel*, 1G S/s each channel Dual channel, 2.5M points on each channel for the Record length;...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 4.Junior User Guidebook This chapter deals with the following topics mainly: Digital Storage Oscilloscope Introduction to the front panel and the user’s interface of the PDS series oscilloscope How to implement the general inspection...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Introduction to the Front Panel and the User's Interface of the PDS series Oscilloscope When you get a new-type oscilloscope, you should get acquainted with its front panel at first and the PDS series mixed digital storage oscilloscope is no exception. This chapter...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 8、 Power and charging indication: Green light indicate AC supply and battery full charged; yellow light indicate under charging. Contrlo(key and knob) area Fig. 4-2 Keys Overview 1、1、Menu option setting: F1~F5 2、Switch...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope User interface introduction Fig. 4-3 Illustrative Drawing of Display Interfaces 1. The Trigger State indicates the following information: Auto: The oscilloscope is under the Automatic mode and is collecting the waveform under the non-trigger state.
Page 14
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 6. It indicates the current function menu. 7. It indicates the operation options for the current function menu, which changes with the function menus. 8. The purple pointer shows the trigger level position.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 22. The positions of two purple dotted line cursors measurements. 23. The reading shows the frequence of the two channels. It is a 6 digits cymometer.Its measurement range of frequency is 2Hz to full bandwidth. when the triggering mode is edge triggering,it is a one channel cymometer and it can only measure the frequency of the tiggering channel.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 10X, Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel. Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten the probe with rotating it to the right side.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope and connect the reference wire clamp with the ground wire connector of the probe connector, and then press the button “7(AUTOSET)”. 2. Check the displayed wave forms and regulate the probe till a correct compensation is achieved (see Fig.4-5 and Fig.4-6).
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.4-7 Attenuation Switch Note: When the attenuation switch is set to 1X, the probe will limit the bandwidth of the oscilloscope in 5MHz. If it is needed to use the whole bandwidth of the oscilloscope, the switch must be set to 10X.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Introduction to the Vertical System Shown as Fig.4-9, there are a series of buttons and knobs in VERTICAL CONTROLS. The following practices will gradually direct you to be familiar with the using of the vertical setting.
“Horizontal Time Base” display in the status bar changes accordingly. The horizontal scanning speed steps from 2 ns up to 100s in the sequence of 1-2-5 -----PDS8102T; 1 ns up to 100s in the sequence of 1-2-5 -----PDS8202T. 2. Use the “HORIZONTAL POSITION” knob to adjust the horizontal position of the signal in the waveform window.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Introduction to the Trigger System Shown as Fig.4-11, there are a knob and four buttons in the “TRIGGER CONTROLS”. The following practices will direct you to be familiar with the setting of the trigger system gradually.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 5.Advanced User Guidebook Up till now, you have already been familiar with the initial operations of the functions of the function areas, buttons and knobs in the front panel of the PDS series oscilloscope.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to Set the Vertical System The VERTICAL CONTROLS includes three menu buttons such as CH1 MENU, CH2 MENU and MATH MENU, and four knobs such as VERTICA POSITION, VOLTS/DIV (one group for each of the two channels).
Page 24
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The description of the Channel Menu is shown as the following list: Function Menu Setting Description Block the DC component in the input signal. Unblock the AC and DC components in the input Coupling signal.
Page 25
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-2 AC Coupling Oscillogram Fig. 5-3 DC Coupling Oscillogram Setting the“Band Limit” Taking the Channel 1 for example, the operation steps are shown as below: (1). Press the CH1 MENU button and call out the CH1 SETUP menu.
Page 26
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Setting the Channel “ON/OFF” Taking the Channel 1 for example, the operation steps are shown as below: (1). Press the CH1 MENU button and call out the CH1 SETUP menu.
Page 27
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-5 Regulation of the Attenuation Ratio of the Probe A List of the Attenuation Coefficient of Probes and the Corresponding Menu Settings. Attenuation Coefficient of the Probe Corresponding Menu Setting...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-6 Wave Form not inverted Fig. 5-7 Wave Form Inverted Implementation Mathematical Manipulation Function The Mathematical Manipulation function is used to show the results of the additive,multiplication,division and subtraction operations between Channel 1 and...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The corresponding FCL (Functional Capabilities List) of the Wave Form Calculation Setting Description CH1-CH2 Subtract the Channel 2 wave form from the Channel 1 wave form. CH2-CH1 Subtract the Channel 1 wave form from the Channel 2 wave form.
Page 30
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope The following table describes the FFT menu: Function Menu Setting Instruction Turn on FFT function. Turn off FFT function. Select CH1 as FFT source. Source Select CH2 as FFT source.
Page 31
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope ■ The FFT feature provides four windows. Each one is a trade-off between frequency resolution and magnitude accuracy. What you want to measure and your source signal characteristics help you to determine which window to use. Use the following guidelines to select the best window.
Page 32
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 1KHz. Fig.5-9. Blackman window Fig.5-10 . Hamming window...
Page 33
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-11. Rectangle window Fig..5-12 . Hanning window Quick Tips If desired, use the zoom feature to magnify the FFT waveform. Use the default dBV RMS scale to see a detailed view of multiple frequencies, even if they have very different amplitudes.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope on the source signal. To reduce random noise and aliased components in repetitive or single-shot events, set the oscilloscope acquisition mode to average. Term interpretation Nyquist frequency: The highest frequency that any Real Time Digital Oscilloscope can measure is exactly half of the sampling rate under the condition of no mistakes,which is called Nyquist frequency.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-13 Information about Vertical Position How to Set the Horizontal system The HORIZONTAL CONTROLS includes the HORIZONTAL NENU button and such knobs as HORIZONTAL POSITION and SEC/DIV. 1. HORIZONTAL POSITION knob: this knob is used to adjust the horizontal positions of all channels (include those obtained from the mathematical manipulation), the analytic resolution of which changes with the time base.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-14 Time Base Mode Menu The description of the Horizontal Menu is as follows: Function Menu Setting Description The setting of the horizontal main time base is Main Time Base used to display the wave form.
Page 37
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-15 Main Time Base Set Window Press the F2 menu selection button and choose Set Window. The screen will show a window area defined by two cursors. In this case, the HORIZONTAL POSITION and SEC/DIV knobs can be used to adjust the horizontal position and size of this window area.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig..5-17 Set Window isable under FFT mode Window Expansion Press the F3 menu selection button and choose Zone Window. As a result, the window area defined by two cursors will be expanded to the full screen size (see Fig.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to set trigger system Trigger is to determine when DSO starts to acquire data and waveform display. Once trigger to be set correctly then it will convert the unstable display to meaning waveform.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-19 Edge trigger menu Edge menu list: MENU SETTING INSTRUCTION Select CH1 as the trigger source. Select CH2 as the trigger source. Source Ext-trigger EXT/5 Ext-trigger divide to 5 and extend trigger level range.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-20 Video trigger menu Video menu list MENU SETTING INSTRUCTION elect CH1 as the trigger source. Select CH2 as the trigger source. Source Ext-trigger EXT/5 Ext-trigger divide to 5 to extend trigger level range...
Page 42
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig. 5-21 Pulse Width Trigger menu Pulse Width Trigger menu list MENU SETTING INSTRUCTION Select CH1 as the trigger source. Source Select CH2 as the trigger source. Mode Pulse (+pulse width less than )
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Slope Trigger Slope trigger sets the oscilloscope as the positive/negative slope trigger within the specified time. Fig. 5-22 The Slope Trigger Menu is shown as Fig. 5-22 Slope Trigge menu...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Alternate trigger Trigger signal comes from two vertical channels when alternate trigger is on. This mode is used to observe two unrelated signals. You can choose different trigger modes for different channels.The options are as follows: edge, video, pulse or slope.
Page 45
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Alternate trigger(Trigger Mode:video) Alternate trigger(Trigger Type:video) Me Fig.5-24 nu is shown as Fig.5-24 Alternate trigger(Trigger Type:video) Menu Alternate trigger(Trigger Type: video) list: Menu MENU SETTING INSTRUCTION elect CH1 as the trigger source.
Page 46
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Alternate trigger(Trigger Mode: Pulse) Alternate trigger(Trigger Type: Pulse) Fig.5-25 Menu is shown as Fig.5-25 Alternate trigger(Trigger Type: Pulse) Menu Alternate trigger(Trigger Type: Pulse) menu list MENU SETTING INSTRUCTION Select CH1 as the trigger source.
Page 47
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Alternate trigger(Trigger Mode: Slope ) Alternate trigger(Trigger Type: Slope )Menu is shown as Fig.5-26. Alternate trigger(Trigger Type: Slope Fig.5-26 )Menu Alternate trigger(Trigger Type: ) menu list: Slope MENU SETTING INSTRUCTION elect CH1 as the trigger source.
Page 48
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Term interpretation 1. Source: Trigger can occur from several sources: Input channels (CH1, CH2), AC Line, Ext, Ext/5. Input :It is the most commonly used trigger source. The channel will work when selected as a trigger source whatever displayed or not.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to Operate the Function Menu The function menu control zone includes 7 function menu buttons and 3 immediate-execution buttons: SAVE/RCL, MEASURE, ACQUIRE, UTILITY, CURSOR, DISPLAY, AUTOSCALE,AUTOSET, RUN/STOP and U-DISK COPY.
Page 50
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-28 Peak Detect mode, under which the burrs on the falling edge of the square wave, can be detected and the noise is heavy. Fig.5-29 Common ACQU Mode display, in which no burr can be detected.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-30 The displayed wave form after the noise is removed under the Average Mode, in which the average number of 16 is set. How to Set the Display System Push down the DISPLAY button and the menu displayed in the screen is shown as Fig.5-31.
Page 52
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Function Menu Setting Description Vectors The space between the adjacent sampling points in the display is filled with the vector form. Type Dots Only the sampling points are displayed. 1sec...
Page 53
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-33 Display in Dots form Persist When the Persist function is used, the persistence display effect of the picture tube oscilloscope can be simulated: the reserved original data is displayed in fade color and the new data is in bright color.
Page 54
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope XY Format This format is only applicable to Channel 1 and Channel 2. After the XY display format is selected, Channel 1 is displayed in the horizontal axis and Channel 2 in the vertical axis;...
Page 55
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-36 Fig.5-36 FFT mode XY Format Disable...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to Save and Recall a Wave Form Press the SAVE/RCL button, you can save and call out the waveforms in the instrument. The menu displayed in the screen is shown as Fig.5-37.
Page 57
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 4. Press the F4 menu selection button and choose ON for CH A. The stored wave form A will be displayed in the screen. The voltage level and time base level will also be shown at the upper left corner of the display area at the same time (see Fig.5-38).
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope How to Implement the Auxiliary System Function Setting Press the UNTILITY button and the menu is displayed in the screen as Fig.5-40. Fig.5-40 Function Menu The description of the Auxiliary Function Menu is shown as the following table.
Page 59
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-41 SYS STAT Menu The “SYS STAT” menu is described as the following table: Function Menu Setting Description Horizontal Show the horizontal parameter of the channel. Vertical Show the vertical parameter of the channel.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Horizontal System State Fig.5-42 How to Implement the Automatic Measurement With the Measure button pressed down, an automatic measurement can be implemented. There are 20 types of measurements and 4 measurement results can be displayed simultaneously.
Page 61
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope +Duty: +Duty Cycle, defined as +Width/Period. -Duty: -Duty Cycle, defined as -Width/Period. Press the F1 menu selection button to choose Source or Type menu. You can choose the channel to be measured from the Source menu and choose the measurement Type (Freq, Period, Mean, PK–PK, and Cyc RMS).
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 9. Press the F4 menu selection button and choose Mean. 10. Press the F5 menu selection button and choose Cyc RMS. The measured value will be displayed in the reading window automatically (see Fig.5-44).
Page 63
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-45 CURS MEAS Menu The description of the cursor measurement menu is shown as the following table: Function Menu Setting Description Switch off the cursor measurement. Voltage Display the voltage measurement cursor and Type menu.
Page 64
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope indicating CURSOR1 and CURSOR2. 3. Press the F2 menu selection button and choose CH1 for Source. 4. Adjust the positions of CURSOR1 and CURSOR2 according to the measured waveform, with the absolute value of the voltage amplitude difference between Cursor 1 and Cursor 2 displayed in the increment window.
Page 65
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.5-47 Wave Form of Cursor Measurement the Cursor Measurement for FFT model: Press the CURSOR button to display the cursor measurement function menu (CURS MEAS) in the screen, which includes Vamp Measurement and Freq Measurement at the mode of FFT, shown as Fig.5-48.
Page 66
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Function Menu Setting Description Switch off the cursor measurement. Vamp Display the Vamp measurement cursor and Type menu. Freq Display the Freq measurement cursor and menu. Source MATH FFT Display the channel for the cursor measure.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Carry out the following operation steps for the Freq cursor measurement : 1. Press CURSOR and recall the CURS MEAS menu. Press F1 and choose Freq for Type, with two purple dotted lines displayed along the vertical direction of the screen indicating the corrersponding Cursor 1 and Cursor 2 3.
Page 68
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Only show one or two periods. Wave Show Multi-period waveforms. If you want to measure the two-channel signal, you can do as the follows: 1. Press Autoscale, the function menu will appear on the right of the screen.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 6. Turn off the submenu at the Autoscale menu, the Autoscale is off and turn on the submenu still enters into the function. 7. When video triggering, the horizontal time scale is 50us. If one channel is showing edge signal, the other channel is showing video one, the time scale refers to 50us as video one as standard.
Page 70
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope corresponding name will be WAVE1.BIN、 WAVE2.BIN、 WAVE3.BIN……or WAVE1.BMP、 WAVE2.BMP、WAVE3.BMP……. Then you can open the files in computer by connecting USB disk to computer. Noted: During the U disk saving procedure, some notes such as “Waveform saving” 、...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 6.Demonstration Example 1: Measurement of Simple Signals Observe an unknown signal in the circuit, and display and measure rapidly the frequency and peak-to-peak value of the signal. Carry out the following operation steps for the rapid display of this signal: 1.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.6-1 Waveform of Automation Measurement Example 2: Gain of the Amplifier in the Metering Circuit Set the probe menu attenuation coefficient as 10X and that of the switch in the probe as 10X.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 11. Calculate the amplifier gain with the following formulas. Gain = Output Signal / Input signal Gain (db) = 20×log (gain) Fig.6-2 Wave Form of Gain Measurement Example 3: Capture the Single Signal The digital storage oscilloscope takes the lead in providing the convenience capturing of such non-periodic signals as pulse and burr, etc.
Page 74
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 6. Press the F1 menu selection button and choose Single as the type. 7. Press the F3 menu selection button and choose Edge as the mode. 8. Press the F4 menu selection button and choose Rising as the slope.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Example 4: Analyze the Details of a Signal Observe the Signal Containing Noises If the signal is interfered by the noise, the noise may cause a failure in the circuit. For...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Fig.6-5 Wave Form of the Noise-Removed Signal Example 5: Application of X-Y Function Examine the Phase Difference between Signals of two Channels Example: Test the phase change of the signal after it passes through a circuit network.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope wave form. 9. With the elliptical oscillogram method adopted, observe and calculate the phase difference (see Fig.6-6). The signal must be centered and kept in the horizontal direction. Fig.6-6 Lissajous Graph Based on the expression sin =A/B or C/D, where, q is the phase difference angle, and the definitions of A, B, C, and D are shown as the graph above.
Page 78
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 2. Press the F1 menu selection button and choose Single for Type. 3. Press the F2 menu selection button and choose CH1 for Source. 4. Press the F3 menu selection button and choose Video for type.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 7.F.A.Q 1. In the case of that the oscilloscope is still in the black-screen state without any display after the power is switch on, implement the following fault treatment procedure.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 8. Technical Specifications Unless otherwise specified, the technical specifications applied are applicable to the probe with the attenuation switch setting 10X and the PDS series digital oscilloscope. Only if the oscilloscope fulfill the following two conditions at first, can these specification standards be reached.
Page 81
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Interpolation (sin x)/x Record length 2.5M points on each channel 2ns/div~100s/div, step by 1~2~5 ---- PDS8102T Scanning speed(S/div) 1ns/div~100s/div, step by 1~2~5 ---- PDS8202T Sampling rate / relay ±100ppm time accuracy Single ±(1...
Page 82
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Vpp, Vmax, Vmin, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot, Preshoot, Freq, Automatic Period, Rise Time,Fall Time, DelayA→B , DelayA→B , +Width, -Width, +Duty, -Duty Waveform Math +, -, *, / ,FFT...
Page 83
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Line number 1-525 (NTSC) and 1-625 (PAL/SECAM) range Positive pulse:>、<、= Trigger condition Slope Trigger negative pulse:>、<、= Time setting 24ns~10s Trigger on CH1 Edge, Pulse, Video, Slope Alternate Trigger Trigger on CH2...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope General Technical Specifications Display Display Type 8” Colored LCD (Liquid Crystal Display) Display Resolution 640 (Horizontal) × 480 (Vertical) Pixels Display Colors 65536 colors, TFT screen Output of the Probe Compensator...
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope 9. Appendix Appendix A: Enclosure Standard Accessories: Passive probe: 2, 1.2 m, 1:1 (10:1) CD: x 1 (PC link application software) RS232 data line or USB data line Power line: one, up to the standards of the country in which it is used.
User Manual of OWON Color Mixed Signal Digital Storage Oscilloscope Appendix C: Battery Using Guide Charging the oscilloscope The lithium battery maybe not be charged when delivery. Please charge the battery for 12 hours to make sure enough power to supply (the oscilloscope should be turned on during charging) to oscilloscope.