2. Clear Memory
Press the Forward arrow button
from the MultiLogPRO.
3. DO
Sensor Calibration
2
If you need to calibrate a DO
enter the DO
calibration screen:
2
_ _ _ _ _
U s e t h e f o l l o w i n g
m e n u t o c a l i b r a t e
t h e D O 2 e l e c t r o d e .
S a l i n i t y : 0 p p t
Use the arrow buttons to select between 0, 5, 10, 15, 20, 25, 30 and 35 ppt and
press the Enter button to confirm and move to calibrate altitude.
Use the arrow buttons to select between 0, 500, 1000, 1500, 2000, 2500, 3000 and
3500 ft and press the Enter button to confirm.
MultiLogPRO will display the calculated upper limit of the calibrated range and will
exit the DO
calibration screen.
2
The new calibration parameters will be saved until the next time you change them.
4. Screen Contrast
Use the arrow buttons to adjust the LCD screen contrast.
Any contrast adjustment will be saved until the next time you change it.
5. Connect to a Computer or to a TI Calculator
Press the Forward arrow
arrow
button to connect to a TI calculator
1.2.10.
Internal Clock and Calendar
The internal clock is set the first time you use the Setup command from the MultiLab
software to program the MultiLogPRO. The clock is automatically updated to the
PC's time and date whenever you connect your MultiLogPRO to a PC and perform
Setup in MultiLab.
if you want to delete all previous data files
sensor press the Forward arrow button
2
D O 2 C A L _ _ _ _ _
button to connect to a computer or the Backward
Chapter 1 MultiLogPRO
to
19