Flow Switch Replacement; Figure 7-50. Hpvs Remove Bd - Thermo Scientific 42iQHL Instruction Manual

High level no-no2-nox analyzer

Advertisement

Servicing
DMC Ozonator
Figure 7–50. HPVS Remove Bd
Flow Switch
Use the following procedure to replace the flow switch.
Replacement
Equipment Required:
Phillips drive, #2
1. Remove Ozonator as described in
"DMC
Ozonator" in this chapter.
2. Unplug flow switch (J5) from board.
3. Pull flow switch off clip after unplugging from board.
Thermo Scientific
42iQHL Instruction Manual
7-49