Self Test

Keysight M9614A/M9615A provides the self test function to check the operation.
It is recommended to perform the self test for the following condition or purpose.
Before performing the self test, turn the instrument output off and disconnect
cables from the measurement terminals.
To perform
See
self test

Self Calibration

Keysight M9614A/M9615A provides the self calibration function to maintain the
measurement performance. If the environmental temperature changes 5 C or
more, perform the self calibration. This is effective for the accurate measurements
by minimizing the effect of thermal drift. The self calibration must be performed
after warming-up of 40 minutes.
Before performing the self calibration, turn the instrument output off and
disconnect cables from the measurement terminals.
To perform
1. Click Start menu > Keysight M960x Source Measure Unit > M960x SFP. The
self calibration
2. From the list on the window, highlight the M9614A/M9615A modules to
3. Confirm that the Access indicator on the M9614A/M9615A front panel turns
Keysight M9614A/M9615A Startup Guide, Edition 1
If emergency shutdown occurs.
In this condition, both the Access and Status indicators turn red. The shutdown
condition will be solved after passing the self test.
If you feel that the instrument may be defective.
For preventive maintenance.
"To Perform Self Test" on page
Connect to Instrument dialog box will open.
connect and click Connect to launch the Keysight M960x Source Measure Unit
Soft Front Panel.
green. Refer to
Figure 1-1
17.
and
Table 1-3
for LED status.
Maintenance
33