Do Not Test Too Quickly; Measurement Signal Error - Proceq PAROTESTER 2 Operating Instructions Manual

Paper and film tester
Table of Contents

Advertisement

10.1 Do not test too quickly!

Fig.64

10.2 Measurement signal error

Fig.65
During the next correctly executed measurement or
when the device problem has been rectified, the values
and current statistics appear again.
24
.
Error messages
.
If the measurement is carried
out too quickly, the impact
body is additionally acceler-
ated.
The message «Measure
more slowly» appears on the
LCD display. (Fig. 64). The
message disappears when
the next correctly executed
measurement is made.
Measurement signal errors
are displayed at the top of
the display instead of the
measurements and statis-
tics. The message «Signal
failure» appears on the LCD
screen with 2 values (Fig.
65) A = X.X Volt and
dt = X.X ms
10.3 Description of the measurement
signal errors
With the impact devices type P, PG and U with error
A = X.XX V
If the A-value is below 3.30 V.
a    C lean the impact bodies and guide tube.
Refer to section 11 «Maintenance»
If the A-value is higher than 4.00 V.
a    R epeat the measurement with a slower loading
movement (section 10.1).
With the impact devices type P, PG and U with error
dt = X.XX ms
If the dt-value is lower than 0.40 ms
The support ring is worn; i.e. the impact body is in
a
an incorrect position at the time of measurement.
Replace the support ring.
On impact device type P with error dt = X.XX ms
If the dt-value is higher than 1.25 ms
The support ring is not correctly positioned.
a
Fit the support ring correctly on the end stop.
On impact device type PG with error dt = X.XX ms
If the dt-value is higher than 1.40 ms
The support ring is not correctly positioned.
a
Fit the support ring correctly on the end stop.
© 2012 Proceq SA

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents