Appendix
Device temperature is be-
coming critical
Device temperature is criti-
cal - Measurements are
disabled
Unknown sensor error
[Measuring program
name]: Missing input data
configuration
[Measuring program
name]: Unknown measur-
ing program state
[Measuring program
name]: Unspecified error
[Measuring program
name]: No input points
[Measuring program
name]: No ROI is defined
[Measuring program
name]: Invalid input object
[Measuring program
name]: Inconsistent num-
ber of objects
[Measuring program
name]: No points in region
of interest
[Measuring program
name]: Too few points in
region of interest
[Measuring program
name]: Too few points are
detected
[Measuring program
name]: The object can't be
fitted (Points are too close
to each other)
[Measuring program
name]: The object can't be
fitted (Collinear points)
The sensor temperature is very
high
The sensor temperature is too
high for performing measure-
ments
An unknown error has occurred
in the sensor.
The input data have not been
configured in the measuring
program specified.
An unknown error has occurred
in the measuring program spec-
ified.
An unspecified error has oc-
curred in the measuring pro-
gram specified.
The input point cloud of the
specified measuring program
contains no points.
The specified measuring pro-
gram requires at least one ROI
in order to run properly.
The input object of the specified
measuring program is invalid.
The number of input objects for
the specified measuring pro-
gram is inconsistent.
There are no measuring points
in the ROI of the measuring pro-
gram specified.
There are not enough points in
the ROI of the specified measur-
ing program for calculating the
measured values.
Not enough points were de-
tected in the specified measur-
ing program for calculating the
measured values
The target object could not be
calculated in the measuring
program specified, as the se-
lected points in the measuring
program are too close together.
The target object could not be
calculated in the measuring
program specified, as the se-
lected points in the measuring
program are collinear.
Wait some time before you start the
next measurement. Decrease the
ambient temperature.
Wait some time before you start the
next measurement. Decrease the
ambient temperature.
Contact Micro-Epsilon.
Configure the input data of the
measuring program (see Chapter
3.16.3)
Contact Micro-Epsilon.
Contact Micro-Epsilon.
Check the parameters of the meas-
uring program generating the point
cloud.
Set the parameters for the ROI in
the measuring program specified
(see Chap. 3.16.4).
Check the parameters of the meas-
uring program generating the input
object.
Contact Micro-Epsilon.
Enlarge the ROI (see Chap. 3.16.4).
Check whether the object being
measured is within the current
measuring range. Correct the set-
tings for the measuring range (see
Chap. 3.4.4). Correct the sensor pa-
rameters (see Chap.
3.4.1
3.4.3).
Enlarge the ROI (see Chap. 3.16.4).
Check whether the object being
measured is within the current
measuring range. Correct the set-
tings for the measuring range (see
Chap. 3.4.4). Correct the sensor pa-
rameters (see Chap.
3.4.1
3.4.3).
Check the parameters of the meas-
uring program specified.
Enlarge the ROI (see Chap. 3.16.4).
Check the parameters of the meas-
uring program specified.
Enlarge the ROI (see Chap. 3.16.4).
Check the parameters of the meas-
uring program specified.
and
and
Page 127