Setting loop graph scaling

Setting patient and sensor type

WARNING

Setting the data source

3-28
Use Spirometry Scaling to set the scales of the spirometry loop
graph. The available settings for the volume, Paw, and flow graph
axis are dependent on the set patient type of adult or pediatric.
Auto automatically adjusts the volume, Paw, and flow axis of the
loop based on the minimum and maximum breath reading
shown in the waveform.
Linked links the adjustment of the volume, Paw, and flow axis of
the loop graph together. Change one of the scales and the
remaining two scales automatically change based on the one
set scale.
Indep. allows the axis of the loop graph to be changed
separately for the volume, Paw, and flow axis.
1.
Select Spirometry - Setup Loops.
2.
Select Spirometry Scaling and set the scale type from the
drop-down list.
3.
Select Back to view changes made and access other functions
of the Spirometry menu.
Patient and sensor type refer to the style of airway adapter used with
the airway module. If spirometry data is obtained from the airway
module, make sure that the sensor type matches the type of airway
adapter used. Adult or pediatric patient types are available.
Make sure that the set sensor type corresponds to the
type of airway adapter in use. If the sensor type is not set
correctly, the information displayed may not be accurate.
1.
Select Spirometry - Setup Loops.
2.
Select Patient and Sensor and then select Adult or Pedi
depending on the sensor used.
3.
Select Back to view changes made and access other functions
of the Spirometry menu.
Use Data Source to specify the source of spirometry data.
2076152-001