Fault
Insufficient cuff pressure
No option to select
Multiple measurement in
the NIBP settings dialog
window
Blood pressure values do
not appear in the EMR
system or in the
PC
seca analytics 115
software
COVIDIEN
FILAC
TM
TM
temperature measurement
Fault
No temperature
measurement possible
Implausible measuring
results,
temperature measurement
unsuccessful
Temperature probe cannot
be pushed completely into
the probe holder
Incorrect blood pressure cuff used
Blood pressure cuff or compressed air
tube leaking
Pump in device faulty
Administrator has configured a
measuring profile. The name of the
measuring profile is displayed.
Multiple measurement was performed
3000
Temperature module of the
SmartBucket not activated
Special patient condition, such as
hypothermia
Rectal measurement performed with
blue temperature probe
Oral/axillary measurement performed
with red temperature probe
Patient activity before oral temperature
measurement:
• Physical exertion
• Eating/drinking
• Brushing teeth
• Smoking
Set measuring position does not match
actual measuring position
No probe cover used
Temperature probe faulty
Probe cover not discarded
Cause
Cause
Remedy
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Dispose of blood pressure cuff, use
replacement
Do not continue using device and have it
repaired by seca Service
Not a malfunction
• Press the name of the measuring profile
• ➔
Starting a multiple
measurement.
For current measurement:
• ➔
and ➔
Modifying presets
single measurement
Administrator:
• Specify Individual measurement as preset
Presets for blood pressure
Remedy
Administrator:
Activate temperature module:
System\SmartBucket tab
• Assess patient vital signs using alternative
means
• Switch from predictive measurement
mode to direct mode
Only perform rectal measurement with red
temperature probe
Only perform oral/axillary measurement
with blue temperature probe
Only perform oral temperature
measurement about 20 minutes after any
of these activities
• Select measuring position to suit the
probe used
• Set correct measuring position on device
• Disinfect the temperature probe
Disinfecting
• Use probe cover
Dispose of temperature probe, use
replacement
• Carefully withdraw temperature probe and
probe cover from the probe holder
• Discard probe cover
• Push the temperature probe into the
probe cover
Troubleshooting • 87
English