8. HYGIENE TREATMENT

8.1 Cleaning

Component (depending on version)
: Monitor with
seca mVSA 535
SmartBucket
Blood pressure cuff and compressed air
tube
Temperature probe (red/blue) with cable
80 •
seca mVSA 535
Cleaning
Disinfecting
Sterilizing
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
► Ensure that the device is switched off before performing any hygiene
treatment.
► Disconnect the power supply connector before performing any
hygiene treatment.
► Before each hygiene treatment, take the rechargeable battery out of
the device (if present and removable).
► Ensure that no fluids penetrate the device.
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
► Use only disinfectants free of chlorine and alcohol which are explic-
itly suitable for acrylic sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium compounds, for example).
► Do not use caustic or abrasive detergents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
► Use disinfectants including the active ingredient 70 % isopropanol
only for measuring accessories for measuring vital signs.
► Clean the device and its accessories as described in the table:
Interval
1. Remove all the measuring accessories (measuring
devices and consumables) from the device (depending
on version)
• ➔
As required
Removing the probe holder
2. Moisten a soft cloth with a soap solution
3. Wipe over all surfaces
4. Allow to air-dry for approx. 30 minutes
1. Moisten a soft cloth with a mild soap solution
2. Clean blood pressure cuff and compressed air tube
As required
3. Rinse thoroughly with water
4. Allow to dry at room temperature
1. Discard probe cover and dispose of it
2. Moisten a soft cloth with a mild soap solution
3. Clean temperature probe
As required
4. Shake out temperature probe so that no liquid remains
in it
5. Allow to air-dry for approx. 30 minutes
Removing/fitting probe holder (devices
with temperature probe)
Cleaning