Seca mVSA 535 Instructions For Use Manual page 45

Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Starting a multiple measurement
► Press the
button on the monitor.
The first part-measurement starts (in this case: 6 part-measurements).
Current cuff pressure is displayed.
The symbols for measurement procedure and measuring mode flash (in
this case: Inflation).
The part-measurement stops automatically as soon as valid blood
pressure values are detected.
The waiting time until the next part-measurement starts.
The values for systolic/diastolic blood pressure SYS/DIA are displayed.
The next part-measurement starts automatically.
Once all part-measurements are present, mean blood pressure (mean
value) is displayed.
You have the following options for continuing:
► Measure more vital signs ➔
► Stop the measurement procedure ➔
procedure
NOTE
• Use the Start button to cancel the mean measurement and continue
or to cancel and re-start.
• If inflation does not deliver a measured value, the device automati-
cally switches to deflation.
• If necessary, the device re-pumps several times during the deflation
(re-pumping: Increasing cuff pressure by approx. 50 mmHg fol-
lowed by incremental releasing of that pressure). If no measured
value is delivered even after several re-pumping steps, the proce-
dure is canceled.
Measuring vital signs
Stopping a measurement
Operation • 45