Vital Signs Measurement; General; Blood Pressure Measurement - Seca mVSA 535 Instructions For Use Manual

Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Fault
Patient and user data can
not be restored manually
Patient and user data
cannot initially be
encrypted after software
update

11.2 Vital signs measurement

General

Fault
Vital signs tab is not
displayed
is being
seca mVSA 535
operated on a wheeled
stand: Measuring results
for vital signs are
implausible

Blood pressure measurement

Fault
Implausible measuring
results
86 •
seca mVSA 535
Cause
USB stick contains several backup
copies
Backup copy on the USB stick has
been renamed
There is not enough memory on the
monitor for initial encryption
General
Blood pressure measurement
COVIDIEN
Manual temperature input
SpO
measurement
2
Cause
USB connecting cable of the
SmartBucket not connected to the
monitor
SmartBucket not activated
• Original seca USB connecting cable
not being used
• USB extension cable also in use
Cause
Excessive patient movement
Incorrect blood pressure cuff used
Blood pressure cuff not applied
correctly
Blood pressure cuff applied to an
extremity to which there is intravenous
access
Ensure that there is only one folder on the
USB stick with the designation
<ddMMyyyy_hhmm>_seca_monitor_b
ackup (if necessary, rename other folders)
Ensure that backup copies are saved in the
folder with the designation
<ddMMyyyy_hhmm>_seca_monitor_b
ackup (if necessary, rename folders)
Initial manual encryption by administrator:
Exporting patient and user data
manually
Resetting the device
Restoring patient and user data
manually
• If necessary, make previous device
settings (for example, WiFi connection)
manually
TM
TM
FILAC
3000 temperature measurement
• Connect the USB connecting cable of the
SmartBucket to the monitor
• Restart the device
Administrator:
Activate SmartBucket: ➔
system components
menu
• Use original seca USB connecting cable
(in scope of delivery for the wheeled stand)
• Do not use a USB extension cable
• Ask the patient to move as little as possi-
ble
• Use correct size of blood pressure cuff
• Only use seca blood pressure cuffs
Correctly apply blood pressure cuff, see
instructions for use for blood pressure cuff
Apply blood pressure cuff to a different
extremity
Remedy
Remedy
Managing
in the administrator
Remedy

Advertisement

Table of Contents
loading

This manual is also suitable for:

535 1220 131535 1200 131535 1210 131535 1010 131