11.4 Connection to the seca analytics 115 software

Fault
Data transmission cannot
be set up between device
and
seca analytics 115
seca patient file not found
when searching for a
patient on the device
No access to seca patient
database of the
PC
seca analytics 115
software
Print function on device
not available
Results report not being
printed
Blood pressure values do
not appear in the
PC
seca analytics 115
software
Manual temperature
entries do not appear in
the
seca analytics 115
software
90 •
seca mVSA 535
Software versions not compatible
No seca patient file set up yet
seca patient file is not assigned to the
user in
seca analytics 115
Windows firewall port-blocking is
active, required ports are blocked
No network connection set up between
the device and the PC on which the
seca analytics 115
installed
The PC on which the
seca analytics 115
installed is not switched on
Automatic synchronization is
deactivated on the device
Software versions not compatible
PC printer not switched on
The PC on which the
seca analytics 115
installed is not switched on
No network connection set up between
device and
seca analytics 115
software
No connection set up between
seca analytics 115
printer
Multiple measurement was performed
Manual temperature entries are not
synchronized with the
PC
seca analytics 115
Cause
PC software is
PC software is
PC software is
PC
PC software and PC
PC software
Remedy
Use a compatible version of
seca analytics 115
Compatibility
Create seca patient file ➔
seca patient file
Check whether the seca patient file can be
assigned to the user in
seca analytics 115
Administrator:
Enable the ports required
Administrator:
Setting up a LAN connection to the
network (stationary operation)
Setting up a WiFi connection (mobile
operation)
Setting up a connection to the seca
analytics 115 PC software
Switch on PC
Administrator:
Exporting patient and user data
manually
Use a compatible version of the
PC software: 1.4 Build
seca analytics 115
6657 or higher
Switch on PC printer
Switch on PC
Administrator: Set up LAN connection
Administrator: Set up connection between
PC software and PC
seca analytics 115
printer
User:
• ➔
Modifying presets
• ➔
Starting a single measurement
Administrator:
• Specify Individual measurement as preset
Presets for blood pressure
Not a malfunction
• Enter values in the
seca analytics 115
software manually
Creating a