Reading Off Pulse Rate - Seca mVSA 535 Instructions For Use Manual

Hide thumbs Also See for mVSA 535:
Table of Contents

Advertisement

Reading off pulse rate

54 •
seca mVSA 535
4. Enter body temperature:
a) Enter the value
b) Press the
The value entered appears in the TEMP field of the Vital signs tab.
You have the following options for continuing:
► Measure more vital signs ➔
► Stop the measurement procedure ➔
procedure
Interrogate pulse rate source
WARNING!
Patient hazard, incorrect measurement
A pulse rate determined on the basis of blood pressure or oxygen
saturation is susceptible to artifacts.
► To obtain an exact value, determine pulse rate by means of ECG or
palpation.
Depending on device configuration, pulse rate is determined on the basis of
blood pressure or oxygen saturation.
Pulse rate is displayed until you stop the measurement procedure
Stopping a measurement
Interrogate pulse rate source
1. Press the PR field.
The PR settings dialog window opens.
The source of the pulse rate (NIBP or SpO
2. Press the Confirm button.
The dialog window closes.
button: Confirm entry
Measuring vital signs
Stopping a measurement
procedure.
) is displayed.