If an error occurs when replacing the SD CARD, the POWER LED is lit in red, so insert a writable SD CARD.
SD CARD replacement must be completed within 10 minutes. If it takes more than 10 minutes to replace
the SD CARD, data is lost.
● Every time the SD CARD is replaced, the _CHG number is added to the file name.
Example: When capturing with the file name "TEST.GBD"
First SD CARD: TEST.GBD
Second SD CARD: TEST_CHG1.GBD
● SD CARD cannot be replaced when ring capturing is enabled.
(4) Operation when an alarm is output
When an alarm is output, the ALARM LED is lit in red. An alarm signal is output from the alarm output
terminal when the LED is lit. If the alarm setting is set to "Hold alarm", the alarm is cleared by pressing and
holding the [CLEAR] key.
SD CARD replacement must be completed within 10 minutes. If it takes more than 10 minutes to replace
the SD CARD, data is lost.
5) Operation when an error occurs
If any error occurs while the GLT400 is operating, the POWER LED is lit in red and the buzzer sounds. You
can check the error details from the GLT400 Setting App. Use the same software to clear the error.
For details of the error, refer to "5.3 List of Error Codes".
If the system goes down due to a hardware failure or firmware malfunction, the buzzer will sound
continuously and the POWER LED and START LED flashes in red alternately.
In this case, you cannot get the error log with GLT400 SETTING APP.
If this happens, turn off the power switch and please contact us.
CHAPTER 3 Operation and Function
3-17
Need help?
Do you have a question about the GLT400 and is the answer not in the manual?
Questions and answers