Zero Test; Performance Test Record - Agilent Technologies E4418B Service Manual

Power meters
Hide thumbs Also See for E4418B:
Table of Contents

Advertisement

Performance Tests

Performance Test Record

Performance Test Record
Model Agilent E4418B/E4419B Power Meter
Tested by ___________________________________________
Serial Number _______________________________________
Date _______________
Test

Zero Test

Instrument Accuracy
3.1623 µW
10 µW
31.6228 µW
100 µW
316.227 µW
1 mW
3.1623 mW
10 mW
31.6228 mW
100 mW
Power Reference
P
meas
Interface Test
Functional Test
Low Range, 30 dBm pad
Low Range, 30 dBm pad
Low Range, no pad
Upper Range, no pad
2-30
Table 2-8: Performance Test Record
Min
Result
−76.40 nW
3.0977 µW
9.904 µW
31.428 µW
99.49 µW
314.72 µW
0.9954 mW
3.1477 mW
9.954 mW
31.477 mW
99.54 mW
0.999 mW
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Channel A
Channel B
Actual
Actual
Result
Result
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
___________
Agilent E4418B/E4419B Service Guide
Max
Result
+76.40 nW
3.2269 µW
10.096 µW
31.819 µW
100.51 µW
317.74 µW
1.0047 mW
3.1769 mW
10.046 mW
31.769 mW
100.46 mW
1.001 mW
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail