[:INPut]:CURRent:AUTO:ELEMent<x>
Function
Turns ON/OFF the current auto range of the
power measurement element or queries the
current setting.
Syntax
[:INPut]:CURRent:AUTO:
ELEMent<x> {<Boolean>}
[:INPut]:CURRent:AUTO:ELEMent<x>?
<x> = 1 to 4 (power measurement element)
Example
:INPUT:CURRENT:AUTO:ELEMENT1 ON
:INPUT:CURRENT:AUTO:ELEMENT1? ->
:INPUT:CURRENT:AUTO:ELEMENT1 1
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
[:INPut]:CURRent:MRANge? (Measured
RANge)
Function
Queries the current range with respect to the
present measured value.
Syntax
[:INPut]:CURRent:MRANge? {<NRf>}
<NRf> = 1 to 4 (power measurement element)
Example
:INPUT:CURRENT:MRANGE? 1 ->
5.00E+00
:INPUT:CURRENT:MRANGE? ->
5.00E+00,5.00E+00,5.00E+00,5.00E+00
Description • Returns the measurement range displayed on
the right side of the screen. The present
measurement range can be retrieved even
when set to auto range.
• If the parameter is omitted, the measurement
ranges of all built-in power measurement
elements are output in order starting with
element 1.
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
[:INPut]:CURRent:RANGe?
Function
Queries the current ranges of all power
measurement elements.
Syntax
[:INPut]:CURRent:RANGe?
Example
:INPUT:CURRENT:RANGE? -> :INPUT:
CURRENT:RANGE:ELEMENT1 5.00E+00;
ELEMENT2 5.00E+00;
ELEMENT3 5.00E+00;
ELEMENT4 5.00E+00
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
IM 760151-17E
[:INPut]:CURRent:RANGe[:ALL]
Function
Collectively sets the current ranges of all power
measurement elements.
Syntax
[:INPut]:CURRent:RANGe
[:ALL] {<Voltage>|<Current>|AUTO}
• For a 5-A input element
<Current> = 10, 20, 50, 100, 200, 500 (mA),
1, 2, or 5 (A) (when TERMinal = DIRect)
<Voltage> = 50, 100, 250, 500 (mV), 1, 2.5,
5, or 10 (V)
(when TERMinal = SENSor)
AUTO = Auto range
• For a 50-A input element
<Current> = 1, 2, 5, 10, 20, or 50 (A)
(when TERMinal = DIRect)
<Voltage> = 50, 100, 250, 500 (mV), 1, 2.5,
5, or 10 (V)
(when TERMinal = SENSor)
AUTO = Auto range
Example
:INPUT:CURRENT:RANGE:ALL 5A
Description • The selectable range is determined by the
input element type (5A/50A) of element 1 and
the current measurement terminal setting
([:INPut]:CURRent:TERMinal:ELEMent1).
Therefore, only the elements that are set to
the same input element type and current
measurement terminal setting as element 1
are set.
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
5.10 INPut Group
5
5-49