* Numeric Data Format
(1) Normal Data
• Phase difference φ (PHI) of a power
measurement element in 180° (Lead/Lag) display
ASCII: "D/G" + <NR3> format (mantissa:
maximum significant digits = 5, exponent: 2
digits, example: G90.00E+00)
FLOAT: IEEE single-precision floating point
(4-byte) format
• Σ of the power value (P, S, Q, PC)
• Integrated value (WH, WHP, WHM, AH, AHP,
AHM)
ASCII: <NR3> format (mantissa: maximum
significant digits = 6, exponent: 2 digits,
example: [-]123.456E+00)
FLOAT: IEEE single-precision floating point
(4-byte) format
• Elapsed time of integration (TIME)
ASCII: <NR1> format in units of seconds
(example: for 1 hour (1:00:00), 3600)
FLOAT: IEEE single-precision floating point
(4-byte) format in units of seconds (example:
for 1 hour (1:00:00), 0x45610000)
• No items (NONE)
ASCII: "NAN" (Not A Number)
FLOAT: 0x7E951BEE(9.91E+37)
• Other than above
ASCII: <NR3> format (mantissa: maximum
significant digits = 5, exponent: 2 digits,
example: [-]123.45E+00)
FLOAT: IEEE single-precision floating point
(4-byte) format
(2) Error Data
• Data does not exist (display: "---------")
ASCII: "NAN" (Not A Number)
FLOAT: 0x7E951BEE(9.91E+37)
• Over the range (display: "---O L---")
• Overflow (display: "---O F---")
• Data over (display: " Error ")
ASCII: "INF" (INFinity)
FLOAT: 0x7E94F56A(9.9E+37)
* List of Numeric Data Output Items That Are Preset
The list of function names used in the commands and
the corresponding function names used on the screen
menu of the WT1600FC is given in the Function
Selection List in the DISPlay group.
Note
The List of Numeric Data Output Items That Are Preset indicates
the measurement function and element that are assigned to each
item number (ITEM<x>). Items that are not set to be measured
are displayed or output in the same fashion as when the data
does not exist. For example, if frequency FI of the current of
element 2 is not set to be measured, the output of item number
ITEM19 is the same as the output when the data does not exist
(NAN for ASCII).
IM 760151-17E
5.13 NUMeric Group
(1)
Preset Pattern of Output Items of Power
Measurement Numeric Data
Applicable command
":NUMeric[:NORMal]:PRESet"
• Pattern 1
<Function>,
ITEM<x>
1
URMS,
2
IRMS,
3
P,
4
S,
5
Q,
6
LAMBda,
7
PHI,
8
FU,
9
FI,
10
NONE
11-19
URMS to FI,
20
NONE
21-29
URMS to FI,
30
NONE
31-39
URMS to FI,
40
NONE
41-49
URMS to FI,
50
NONE
51-59
URMS to FI,
60
NONE
61-69
URMS to FI,
70
NONE
71-79
URMS to FI,
80
NONE
81-89
URMS to FI,
90
NONE
91-255
NONE
• Pattern 2
ITEM<x>
<Function>,
1
URMS,
2
UMN,
3
UDC,
4
UAC,
5
IRMS,
6
IMN,
7
IDC,
8
IAC,
9
P,
10
S,
11
Q,
12
LAMBda,
13
PHI,
14
FU,
15
FI,
16-30
URMS to FI,
31-45
URMS to FI,
46-60
URMS to FI,
61-75
URMS to FI,
76-90
URMS to FI,
91-105
URMS to FI,
106-120
URMS to FI,
121-135
URMS to FI,
136-255
NONE
<Element>
1
1
1
1
1
1
1
1
1
2
5
3
4
5
6
SIGMA
SIGMB
SIGMC
<Element>
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
2
3
4
5
6
SIGMA
SIGMB
SIGMC