YOKOGAWA WT1600FC User Manual page 88

Impedance meter communication interface
Table of Contents

Advertisement

5.10 INPut Group
[:INPut]:MODUle?
Function
Queries the input element type.
Syntax
[:INPut]:MODUle? {<NRf>}
[:INPut]:MODUle?
<NRf> = 1 to 6 (element)
Example
:INPUT:MODULE? 1 -> 5
:INPUT:MODULE? -> 5,5,5,5,Z5,0
Description • The response information is as follows:
5 = 5-A input element for power measurement
50 = 50-A input element for power
measurement
Z5 = 5-A input element for impedance
measurement
Z20 = 20-A input element for impedance
measurement
0 = No input element
• If the parameter is omitted, the input element
types of all elements are output in order
starting with element 1.
[:INPut]:NULL
Function
Turns ON/OFF the NULL function or queries the
current setting.
Syntax
[:INPut]:NULL {<Boolean>}
[:INPut]:NULL?
Example
:INPUT:NULL ON
:INPUT:NULL? -> :INPUT:NULL 1
[:INPut]:POVer?
Function
Queries the peak over information.
Syntax
[:INPut]:POVer?
Example
:INPUT:POVER? -> 0
Description • The peak over information of each element is
mapped as shown below. For the response, a
sum of decimal values of each bit is returned.
• If the response is "16," for example, peak
over is occurring at U3.
15 14 13 12 11 10 9
0
0
0
0
[:INPut]:SCALing?
Function
Queries all settings related to scaling.
Syntax
[:INPut]:SCALing?
Example
:INPUT:SCALING? -> :INPUT:SCALING:
STATE:ELEMENT1 0;ELEMENT2 0;
ELEMENT3 0;ELEMENT4 0;ELEMENT5 0;:
INPUT:SCALING:PT:ELEMENT1 1.0000;
ELEMENT2 1.0000;ELEMENT3 1.0000;
ELEMENT4 1.0000;ELEMENT5 1.0000;:
INPUT:SCALING:CT:ELEMENT1 1.0000;
ELEMENT2 1.0000;ELEMENT3 1.0000;
ELEMENT4 1.0000;ELEMENT5 1.0000;:
INPUT:SCALING:SFACTOR:
ELEMENT1 1.0000;ELEMENT2 1.0000;
ELEMENT3 1.0000;ELEMENT4 1.0000
5-52
8
7
6
5
4
3
2
0
0 I5 U5 I4 U4 I3 U3 I2 U2 I1 U1
[:INPut]:SCALing:{PT|CT|SFACtor}?
Function
Queries the {Voltage|Current|Power} scaling
constants of all elements.
Syntax
[:INPut]:SCALing:{PT|CT|SFACtor}?
Example
:INPUT:SCALING:PT? -> :INPUT:
SCALING:PT:ELEMENT1 1.0000;
ELEMENT2 1.0000;ELEMENT3 1.0000;
ELEMENT4 1.0000;ELEMENT5 1.0000
Description The [:INPut]:SCALing:SFACtor.. commands
are invalid on models that are only equipped with
impedance measurement elements.
[:INPut]:SCALing:{PT|CT|SFACtor}[:ALL]
Function
Collectively sets the {Voltage|Current|Power}
scaling constants of all elements.
Syntax
[:INPut]:SCALing:{PT|CT|SFACtor}
[:ALL] {<NRf>}
<NRf> = 0.0001 to 99999.9999
Example
:INPUT:SCALING:PT:ALL 1
Description • The
• The [:INPut]:SCALing:SFACtor..
[:INPut]:SCALing:{PT|CT|SFACtor}:
ELEMent<x>
Function
Sets the {Voltage|Current|Power} scaling
1
0
constants of the element or queries the current
setting.
Syntax
[:INPut]:SCALing:{PT|CT|SFACtor}:
ELEMent<x> {<NRf>}
[:INPut]:SCALing:{PT|CT|SFACtor}:
ELEMent<x>?
• For [:INPut]:SCALing:{PT|CT}:
• For [:INPut]:SCALing:{SFACtor}:
<NRf> = 0.0001 to 99999.9999
Example
:INPUT:SCALING:PT:ELEMENT1 1
:INPUT:SCALING:PT:ELEMENT1? ->
:INPUT:SCALING:PT:ELEMENT1 1.0000
Description The [:INPut]:SCALing:SFACtor.. commands
are invalid on models that are only equipped with
impedance measurement elements.
"[:INPut]:SCALing:{PT|CT}[:ALL]"
command collectively sets the scaling
constants on all power measurement
elements. To set the scaling constants on
impedance measurement elements, use the
"[:INPut]:SCALing:{PT|CT}:ELEMent<x>"
command. This command can be used only
on models that are equipped with power
measurement elements.
commands are invalid on models that are only
equipped with impedance measurement
elements.
ELEMent<x>
<x> = 1 to 5 (element)
ELEMent<x>
<x> = 1 to 4 (power measurement element)
IM 760151-17E