YOKOGAWA WT1600FC User Manual page 112

Impedance meter communication interface
Table of Contents

Advertisement

5.16 STORe Group
:STORe:FILE:NAME
Function
Sets the name of the file when saving the stored
data or queries the current setting.
Syntax
:STORe:FILE:NAME {<Filename>}
:STORe:FILE:NAME?
Example
:STORE:FILE:NAME "DATA1"
:STORE:FILE:NAME? -> :STORE:FILE:
NAME "DATA1"
Description The save destination of the stored data is
specified using:
• the ":FILE:DRIVe" command for the drive.
• the ":FILE:CDIRectory" command for the
directory.
The save destination path can be queried using
the ":FILE:PATH?" command.
:STORe:INTerval
Function
Sets the store interval or queries the current
setting.
Syntax
:STORe:INTerval {<NRf>,<NRf>,<NRf>}
:STORe:INTerval?
1st <NRf> = 0 to 99 (hour)
2nd <NRf> = 0 to 59 (minute)
3rd <NRf> = 1 to 59 (second)
Example
:STORE:INTERVAL 0,0,0
:STORE:INTERVAL? -> :STORE:
INTERVAL 0,0,0
:STORe:ITEM
Function
Sets the items to be stored or queries the
current setting.
Syntax
:STORe:ITEM {NUMeric|WAVE|NWAVe}
:STORe:ITEM?
NUMeric = Store only the numeric values.
WAVE = Store only the waveforms.
NWAVe = Store both the numeric values and
the waveforms.
Example
:STORE:ITEM NUMERIC
:STORE:ITEM? -> :STORE:ITEM NUMERIC
:STORe:MEMory:CONVert:ABORt
Function
Abort converting the stored data from the
memory to the file.
Syntax
:STORe:MEMory:CONVert:ABORt
Example
:STORE:MEMORY:CONVERT:ABORT
5-76
:STORe:MEMory:CONVert:EXECute
Function
Executes the converting of the stored data from
the memory to the file.
Syntax
:STORe:MEMory:CONVert:EXECute
Example
:STORE:MEMORY:CONVERT:EXECUTE
Description • The convert destination file is set using the
":STORe:FILE: ..." command.
• When file conversion is executed, the
WT1600FC accesses the file twice.
To confirm the completion of the file
conversion, use the "COMMUNICATE:WAIT
64" command (checks the change in bit 6
(ACS) of the condition register) and check the
completion of the file access of the
WT1600FC twice. An example is indicated
below.
"STATUS:EESR?"
(Clear the extended event register)
"STORE:MEMORY:CONVERT:EXECUTE"
(Start the file conversion)
"COMMUNICATE:WAIT 64"
(Wait for the conversion to finish, the first
time)
"STATUS:EESR?"
(Clear the extended event register)
"COMMUNICATE:WAIT 64"
(Wait for the conversion to finish, the
second time)
"STATUS:EESR?"
(Clear the extended event register)
:STORe:MEMory:INITialize
Function
Executes the initialization of the storage
memory.
Syntax
:STORe:MEMory:INITialize
Example
:STORE:MEMORY:INITIALIZE
:STORe:MODE
Function
Sets the data storage/recall or queries the
current setting.
Syntax
:STORe:MODE {STORe|RECall}
:STORe:MODE?
Example
:STORE:MODE STORE
:STORE:MODE? -> :STORE:MODE STORE
IM 760151-17E

Advertisement

Table of Contents
loading

Table of Contents