<Space>
?
MHOLd
<Space>
?
?
:MEASure?
Function
Queries all settings related to the measurement.
Syntax
:MEASure?
Example
• Example for power measurement
:MEASURE? -> :MEASURE:AVERAGING:
STATE 0;TYPE EXPONENT;COUNT 2;:
MEASURE:FREQUENCY:ITEM U1,I1,U2;:
MEASURE:FUNCTION1:STATE 0;
EXPRESSION "URMS(E1)";UNIT "V";:
MEASURE:FUNCTION2:STATE 0;
EXPRESSION "IRMS(E1)";UNIT "A";:
MEASURE:FUNCTION3:STATE 0;
EXPRESSION "UPPK(E1)";UNIT "V";:
MEASURE:FUNCTION4:STATE 0;
EXPRESSION "IPPK(E1)";UNIT "A";:
MEASURE:DMEASURE:OBJECT SIGMA;
TYPE OFF;:MEASURE:SFORMULA RMS;
PC:IEC 1976;P1 0.5000;P2 0.5000;:
MEASURE:PHASE 180;MHOLD 0
• For impedance measurement
:MEASURE? -> :MEASURE:AVERAGING:
STATE 0
:MEASure:AVERaging?
Function
Queries all settings related to averaging.
Syntax
:MEASure:AVERaging?
Example
• Example for power measurement
:MEASURE:AVERAGING? -> :MEASURE:
AVERAGING:STATE 1;TYPE EXPONENT;
COUNT 2
• For impedance measurement
:MEASURE:AVERAGING? -> :MEASURE:
AVERAGING:STATE 1
IM 760151-17E
<NRf>
OFF
ON
<NRf>
:MEASure:AVERaging:COUNt
Function
Sets the averaging coefficient for power
measurement or queries the current setting.
Syntax
:MEASure:AVERaging:COUNt {<NRf>}
:MEASure:AVERaging:COUNt?
<NRf> = 2, 4, 8, 16, 32, or 64
(when TYPE = EXPonent)
<NRf> = 8, 16, 32, 64, 128, or 256
TYPE = LINear)
Example
:MEASURE:AVERAGING:COUNT 2
:MEASURE:AVERAGING:COUNT? ->
:MEASURE:AVERAGING:COUNT 2
Description • This command is valid only during power
measurement. (It cannot be used during
impedance measurement.)
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
• For details on the averaging coefficient
(attenuation constant) during impedance
measurement, see the WT1600FC User's
Manual (IM760151-01E).
:MEASure:AVERaging[:STATe]
Function
Turns ON/OFF averaging or queries the current
setting.
Syntax
:MEASure:AVERaging
[:STATe] {<Boolean>}
:MEASure:AVERaging:STATe?
Example
:MEASURE:AVERAGING:STATE ON
:MEASURE:AVERAGING:STATE? ->
:MEASURE:AVERAGING:STATE 1
Description The averaging for impedance measurement can
only be turned ON/OFF. For details on the
averaging during impedance measurement, see
the WT1600FC User's Manual (IM760151-01E).