YOKOGAWA WT1600FC User Manual page 47

Impedance meter communication interface
Table of Contents

Advertisement

Command
Common Command Group
*CAL?
*CLS
*ESE
*ESR?
*IDN?
*OPC
*OPC?
*OPT?
*PSC
*RST
*SRE
*STB?
*TRG
*TST?
*WAI
IM 760151-17E
Function
Executes zero calibration (zero level compensation, same operation as
pressing CAL (SHIFT+MEASURE)) and queries the result.
Clears the standard event register, extended event register, and error queue.
Sets the standard event enable register or queries the current setting.
Queries the standard event register.
Queries the instrument model.
Sets a "1" to bit 0 (OPC bit) of the standard event register upon the
completion of the specified overlap command.
The register is cleared when the value rounded to an integer is a
non-zero value.
ASCII code "1" is returned when the specified overlap command is
completed when OPC? is transmitted.
Queries the installed options.
Sets whether or not to clear the registers at power on or queries the
current setting.
Executes the initialization of settings.
Sets the service request enable register or queries the current setting.
Queries the status byte register.
Executes the same operation as when SINGLE (SHIFT+HOLD) is pressed.
Performs a self-test and queries the result.
Holds the subsequent command until the completion of the specified
overlap operation.
5.1 Command List
Page
5-87
5-87
5-87
5-88
5-88
5-88
5-88
5-88
5-88
5
5-88
5-88
5-89
5-89
5-89
5-89
5-89
5-11

Advertisement

Table of Contents
loading