YOKOGAWA WT1600FC User Manual page 59

Impedance meter communication interface
Table of Contents

Advertisement

• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
:DISPlay[:NUMeric]:NORMal:ITEM<x>
Function
Sets the numeric display item for power
measurement or queries the current setting.
Syntax
:DISPlay[:NUMeric]:NORMal:ITEM<x>
{NONE|<Function>,<Element>}
:DISPlay[:NUMeric]:NORMal:ITEM<x>?
<x> = 1 to 100 (item number)
NONE = No display item
<Function> = {URMS|UMN|UDC|UAC|IRMS|
...} (See the function selection
list (1).")
<Element> =
{<NRf>|SIGMA|SIGMB|SIGMC}(<NRf> = 1 to 4
(power measurement element))
Example
:DISPLAY:NUMERIC:NORMAL:
ITEM1 URMS,1
:DISPLAY:NUMERIC:NORMAL:ITEM1? ->
:DISPLAY:NUMERIC:NORMAL:
ITEM1 URMS,1
Description • This command is valid when the display
format of numeric values
(:DISPlay[:NUMeric]
:NORMal:IAMount) is set to "<NRf> (split
display)."
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
:DISPlay[:NUMeric]:NORMal:PRESet
Function
Presets the display order pattern of numeric
display items for power measurement.
Syntax
:DISPlay[:NUMeric]:NORMal:
PRESet {<NRf>}
<NRf> = 1 to 4 (pattern number)
Example
:DISPLAY:NUMERIC:NORMAL:PRESET 1
Description • Regardless of what value (1 to 4) is specified
for <NRf>, the display pattern (order) of the
numeric display items will be the same as the
display order when Reset List Exec of the
Display setting menu, which is displayed on
the WT1600FC screen, is executed. For
details on the order of displayed items when
reset is executed, see the WT1600FC User's
Manual (IM760151-01E).
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
IM 760151-17E
:DISPlay:TRENd?
Function
Queries all settings related to the trend.
Syntax
:DISPlay:TRENd?
Example
:DISPLAY:TREND? -> :DISPLAY:TREND:
SAMPLING 1;T1 1;T2 1;T3 1;T4 1;
T5 1;T6 1;T7 1;T8 1;T9 0;T10 0;
T11 0;T12 0;T13 0;T14 0;T15 0;
T16 0;FORMAT SINGLE;TDIV 0,0,3;
NORMAL:ITEM1:FUNCTION URMS,1;
SCALING:MODE AUTO;VALUE 100.00E+00,
-100.00E+00;:DISPLAY:TREND:NORMAL:
ITEM2:FUNCTION IRMS,1;SCALING:
MODE AUTO;VALUE 100.00E+00,
-100.00E+00;...(omitted)...;
:DISPLAY:TREND:NORMAL:ITEM16:
FUNCTION FU,2;SCALING:
MODE AUTO;VALUE 100.00E+00,
-100.00E+00
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
:DISPlay:TRENd:ALL
Function
Collectively turns ON/OFF all trends.
Syntax
:DISPlay:TRENd:ALL {<Boolean>}
Example
:DISPLAY:TREND:ALL ON
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
:DISPlay:TRENd:FORMat
Function
Sets the display format of the trend or queries
the current setting.
Syntax
:DISPlay:TRENd:FORMat {SINGle|DUAL|
TRIad|QUAD}
:DISPlay:TRENd:FORMat?
Example
:DISPLAY:TREND:FORMAT SINGLE
:DISPLAY:TREND:FORMAT? ->
:DISPLAY:TREND:FORMAT SINGLE
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
5.4 DISPlay Group
5
5-23