YOKOGAWA WT1600FC User Manual page 61

Impedance meter communication interface
Table of Contents

Advertisement

Description • Set the upper limit and then the lower limit.
• This command is valid when the scaling
mode of the trend
(:DISPlay:TRENd:NORMal:ITEM<x>:
SCALing:MODE) is set to "MANual."
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
:DISPlay:TRENd:PDIV
Function
Sets the horizontal axis (Point/div) of the trend
or queries the current setting.
Syntax
:DISPlay:TRENd:PDIV {<NRf>}
:DISPlay:TRENd:PDIV?
<NRf> = 1, 2, 5, 10, 20, 50, 100, 200, or 500
Example
:DISPLAY:TREND:PDIV 50
:DISPLAY:TREND:PDIV? -> :DISPLAY:
TREND:PDIV 50
Description • This command is valid when waveform
sampling (:WSETup[:SAMPling]) is set to
ON during power measurement.
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
:DISPlay:TRENd:RESTart
Function
Restarts the trend.
Syntax
:DISPlay:TRENd:RESTart
Example
:DISPLAY:TREND:RESTART
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
:DISPlay:TRENd[:SAMPling]
Function
Turns ON/OFF the trend waveform sampling or
queries the current setting.
Syntax
:DISPlay:TRENd:
[:SAMPling] {<Boolean>}
:DISPlay:TRENd:[:SAMPling]?
Example
:DISPLAY:TREND:SAMPLING ON
:DISPLAY:TREND:SAMPLING? ->
:DISPLAY:TREND:SAMPLING 1
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
IM 760151-17E
:DISPlay:TRENd:TDIV
Function
Sets the horizontal axis (T/div) of the trend for
power measurement or queries the current
setting.
Syntax
:DISPlay:TRENd:TDIV {<NRf>,<NRf>,
<NRf>}
:DISPlay:TRENd:TDIV?
{<NRf>, <NRf>, <NRf>} = 0, 0, 3 to 24, 0, 0
1st <NRf> = 1, 3, 6, 12, or 24
2nd <NRf> = 2, 3, 6, 10, or 30 (minute)
3rd <NRf> = 3, 6, 10, or 30
Example
:DISPLAY:TREND:TDIV 0,0,3
:DISPLAY:TREND:TDIV? -> :DISPLAY:
TREND:TDIV 0,0,3
Description • Set the three <NRf>'s so that one <NRf> is a
non-zero value and the other two are zeroes.
• This command is valid when waveform
sampling (:WSETup[:SAMPling]) is set to
OFF.
• This command is invalid on models that are
only equipped with impedance measurement
elements, since the mode is fixed to
impedance measurement.
:DISPlay:TRENd:T<x>
Function
Turns ON/OFF the trend or queries the current
setting.
Syntax
:DISPlay:TRENd:T<x> {<Boolean>}
:DISPlay:TRENd:T<x>?
<x> = 1 to 16 (item number)
Example
:DISPLAY:TREND:T1 ON
:DISPLAY:TREND:T1? ->
:DISPLAY:TREND:T1 1
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
:DISPlay:WAVE?
Function
Queries all settings related to the waveform
display.
Syntax
:DISPlay:WAVE?
Example
:DISPLAY:WAVE? -> :DISPLAY:WAVE:
U1 1;U2 1;U3 1;U4 1;I1 1;I2 1;I3 1;
I4 1;FORMAT SINGLE;
INTERPOLATE LINE;GRATICULE GRID;
SVALUE 1;TLABEL 0;MAPPING:MODE AUTO
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
5.4 DISPlay Group