YOKOGAWA WT1600FC User Manual page 121

Impedance meter communication interface
Table of Contents

Advertisement

:WSETup?
Function
Queries all settings related to the waveform
observation.
Syntax
:WSETup?
Example
:WSETUP? -> :WSETUP:SAMPLING 0;
TDIV 500.0E-06;TRIGGER:MODE AUTO;
SOURCE U1;SLOPE RISE;LEVEL 0.0;:
WSETUP:VZOOM:U1 1.00;U2 1.00;
U3 1.00;U4 1.00;I1 1.00;I2 1.00;
I3 1.00;I4 1.00;:WSETUP:POSITION:
U1 0.000;U2 0.000;U3 0.000;
U4 0.000;I1 0.000;I2 0.000;
I3 0.000;I4 0.000;
:WSETup:POSition?
Function
Queries all settings related to the vertical
position (GND position) of the waveform.
Syntax
:WSETup:POSition?
Example
:WSETUP:POSITION? -> :WSETUP:
POSITION:U1 0.000;U2 0.000;
U3 0.000;U4 0.000;I1 0.000;
I2 0.000;I3 0.000;I4 0.000;
:WSETup:POSition:{UALL|IALL}
Function
Collectively sets the vertical position (level of
the center position) of the waveform
{voltage|current} of all elements.
Syntax
:WSETup:POSition:{UALL|
IALL} {<NRf>}
<NRf> = –130.000 to 130.000(%)
Example
:WSETUP:POSITION:UALL 0
:WSETup:POSition:{U<x>|I<x>}
Function
Sets the vertical position (level of the center
position) of the waveform {voltage|current} of
the power measurement element or queries the
current setting.
Syntax
:WSETup:POSition:{U<x>|
I<x>} {<NRf>}
:WSETup:POSition:{U<x>|I<x>}?
<x> = 1 to 4 (power measurement element)
<NRf> = –130.000 to 130.000(%)
Example
:WSETUP:POSITION:U1 0
:WSETUP:POSITION:U1? -> :WSETUP:
POSITION:U1 0.000
:WSETup[:SAMPling]
Function
Turns ON/OFF the waveform sampling or
queries the current setting.
Syntax
:WSETup[:SAMPling] {<Boolean>}
:WSETup:SAMPling?
Example
:WSETUP:SAMPLING ON
:WSETUP:SAMPLING? -> :WSETUP:
SAMPLING 1
IM 760151-17E
5.19 WSETup (Wave SETup) Group
:WSETup:TDIV
Function
Sets the Time/div value of the waveform or
queries the current setting.
Syntax
:WSETup:TDIV {<Time>}
:WSETup:TDIV?
<Time> = 0.5, 1, 2, 5, 10, 20, 50, 100, 200, 500
(ms)
Example
:WSETUP:TDIV 0.5MS
:WSETUP:TDIV? -> :WSETUP:
TDIV 500.0E-06
Description The specifiable Time/div value is up to 1/10 of
the data update rate (:RATE).
:WSETup:TRIGger?
Function
Queries all settings related to the trigger.
Syntax
:WSETup:TRIGger?
Example
:WSETUP:TRIGGER? ->
:WSETUP:TRIGGER:MODE AUTO;
SOURCE U1;SLOPE RISE;LEVEL 0.0
:WSETup:TRIGger:LEVel
Function
Sets the trigger level or queries the current
setting.
Syntax
:WSETup:TRIGger:LEVel {<NRf>}
:WSETup:TRIGger:LEVel?
<NRf> = –100.0 to 100.0 (%) (The resolution is
0.1(%))
Example
:WSETUP:TRIGGER:LEVEL 0
:WSETUP:TRIGGER:LEVEL? -> :WSETUP:
TRIGGER:LEVEL 0.0
Description Set the value in terms of a percentage of the full
scale value displayed on the screen.
:WSETup:TRIGger:MODE
Function
Sets the trigger mode or queries the current
setting.
Syntax
:WSETup:TRIGger:MODE {AUTO|NORMal}
:WSETup:TRIGger:MODE?
Example
:WSETUP:TRIGGER:MODE AUTO
:WSETUP:TRIGGER:MODE? -> :WSETUP:
TRIGGER:MODE AUTO
:WSETup:TRIGger:SLOPe
Function
Sets the trigger slope or queries the current
setting.
Syntax
:WSETup:TRIGger:SLOPe {RISE|FALL|
BOTH}
:WSETup:TRIGger:SLOPe?
Example
:WSETUP:TRIGGER:SLOPE RISE
:WSETUP:TRIGGER:SLOPE? -> :WSETUP:
TRIGGER:SLOPE RISE
5