YOKOGAWA WT1600FC User Manual page 53

Impedance meter communication interface
Table of Contents

Advertisement

:CURSor:WAVE[:STATe]
Function
Turns ON/OFF the cursor display on the
waveform display or queries the current setting.
Syntax
:CURSor:WAVE[:STATe] {<Boolean>}
:CURSor:WAVE:STATe?
Example
:CURSOR:WAVE:STATE ON
:CURSOR:WAVE:STATE? -> :CURSOR:
WAVE:STATE 1
:CURSor:WAVE:TRACe<x>
Function
Sets the cursor target on the waveform display
or queries the current setting.
Syntax
:CURSor:WAVE:TRACe<x> {U<x>|I<x>}
:CURSor:WAVE:TRACe<x>?
<x> of TRACe<x> = 1 and 2
<x> of U<x>, I<x> = 1 to 4 (power measurement
element)
Example
:CURSOR:WAVE:TRACE1 U1
:CURSOR:WAVE:TRACE1? ->
:CURSOR:WAVE:TRACE1 U1
:CURSor:WAVE:{X<x>|DX|PERDt|Y<x>|DY}?
Function
Queries the cursor measurement value on the
waveform display.
Syntax
:CURSor:WAVE:{X<x>|DX|PERDt|Y<x>|
DY}?
X<x> = X-axis value of the cursor position
(X1=X+, X2=Xx)
DX = X-axis value between cursors (∆X)
PERDt = 1/∆T (1/∆X) value between cursors
Y<x> = Y-axis value of the cursor position
(Y1=Y+, Y2=Yx)
DY = Y-axis value between cursors (∆Y)
<x> = 1, 2
Example
:CURSOR:WAVE:Y1? -> 78.628E+00
Description If the cursor display is not turned ON in the
waveform display, "NAN (Not A Number)" is
returned.
IM 760151-17E
5.3 CURSor Group
5
5-17